Metrohm IC Net 2.3 Manual de usuario Pagina 83

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 389
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 82
6.3 732 IC Detector
IC Net 2.3
73
List of program instructions
The following program instructions can be added to the time pro-
gram on the Program subpage:
Instruction Parameter entry Meaning
Remote 0, 1, *
Set remote output lines 1...8 to
the desired values. For entry of
the first value, enter 1, 0 or *. For
entry of the other values, move
the cursor in front of the value to
be changed and enter
1, 0 or *.
Flag reset, end Set program flag. The end flag
can be used to end a program,
especially if the program time
should be longer than the chro-
matogram duration. Additional
steps after this flag are not al-
lowed. The reset flag is used to
reset the parameters to the sys-
tem startup values.
Range 100, 200, 5000 µS/cm
1, 2, 5, 10 mS/cm
Set measuring range to the se-
lected value.
FS 0.05 µS/cm ...
10 mS/cm
Set full scale range to the se-
lected value (depends on Range
setting).
Zero on, off Switch on or off the auto-zero
function.
Polarity +, Reset the polarity of the output
signal.
Mark Trigger marking signal.
ValveA fill, inject Switch injection valve A to "inject"
or "fill" position.
ValveB fill, inject Switch injection valve B to "inject"
or "fill" position.
Suppressor Switch suppressor module to the
next position.
PumpRS on, off Switch on or off the high-pressure
pump connected to the 732 IC
Detector.
Flow 0.01 ... 5 mL/min Set flow rate of the high-pressure
pump to the desired value.
Vista de pagina 82
1 2 ... 78 79 80 81 82 83 84 85 86 87 88 ... 388 389

Comentarios a estos manuales

Sin comentarios