Metrohm IC Net 2.3 Manual de usuario Pagina 145

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 389
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 144
6.12 Metrohm 709 Solvent Delivery Unit
IC Net 2.3
135
Manual
The Manual tab of the Pump Control Unit window is only available
for a connected system.
Pressure Live display of current pressure in MPa.
This field is also available for the WATCH
WINDOW
.
Flow Live display of current flow in mL/min. This
field is also available for the WATCH WIN-
DOW
.
New flow A new flow can be set by moving the scroll
bar, by entering a value or by changing the
value using the up and down arrows.
Entry range:
0.01 ... 20.00 mL/min (depending
on number of pumps)
Min/Max pressure The minimum and maximum pressure limit
for the individual pump units can be set in
steps of 0.1 MPa.
Entry range: 0.0 ... 50.0 MPa.
%B Individual flow rate for pump B in percent of
total flow.
Entry range:
0 ... 100 %
%C Individual flow rate for pump C in percent of
total flow.
Entry range: 0 ... 100 %
%D Individual flow rate for pump D in percent of
total flow.
Entry range: 0 ... 100 %
Vista de pagina 144
1 2 ... 140 141 142 143 144 145 146 147 148 149 150 ... 388 389

Comentarios a estos manuales

Sin comentarios