Metrohm IC Net 2.3 Manual de usuario Pagina 218

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 389
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 217
6 Devices
IC Net 2.3
208
Entry range: 0…999 µL, default: 30 (with air
segment)
The minimum flush volume depends on
the volume of needle and needle tubing,
flush volumes of less than twice the vol-
ume of needle and tubing will result in
decreased performance.
Inj. volume, µL Programmable only for Partial and Pick-up
modes, is equal to the
Loop volume in
Flushed mode. Is loaded from the sample
queue. The maximum injection volume de-
pends on the injection mode:
Partial: 50% of the loop volume
Pick-up: (loop volume – 3 × needle vol
ume)/2
Entry range: 1…1000 µL
Analysis time Time between switching the injection valve
to inject and the start of processing the next
sample. Is loaded from the sample queue.
Entry range: 0:00:00…9:59:59 h:mm:ss
Delay time Time between start determination and start
of the Triathlon methods/User program, useful
e.g. for baseline equilibrium after a sup-
pressor step.
Entry range: 0:00:00…1:00:00 h:mm:ss
<Set> Apply parameter values selected under
Method values to the Injection method.
Vista de pagina 217
1 2 ... 213 214 215 216 217 218 219 220 221 222 223 ... 388 389

Comentarios a estos manuales

Sin comentarios