6.13 Metrohm 818 Solvent Delivery Unit
IC Net 2.3
141
Stop after Stop time for all pump drives.
Entry range: 0 ... 99999 min
<Set> Send current parameters immediately to the
pump control system. Parameters are not
stored in the system file (*.smt) as long as
the file is not saved.
<Start> Start all pump drives.
<Stop> Stop all pump drives.
Program
The
Program tab of the Pump Control Unit window contains a time
program for the pump control system to define eluent gradients.
Flow Overall flow of pump control system in
mL/min. This value is also available for the
WATCH WINDOW. This field is only available
if the Show flow column option is disabled.
Entry range: 0.01 ... 20.00 mL/min (depending
on number of pumps)
Stop pumps after the run
Stop pumps after the determination.
Program table
Number Row number in the program table (read-
only).
Time Time at which the flow and/or gradient is
changed.
Entry range:
0.0 ... 99999 min