Metrohm IC Net 2.3 Manual de usuario

Busca en linea o descarga Manual de usuario para Software Metrohm IC Net 2.3. Metrohm IC Net 2.3 User Manual Manual de usuario

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 389
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente

Indice de contenidos

Pagina 1 - IC Net 2.3 SR5

CH-9101 Herisau/Switzerland Internet www.metrohm.com E-Mail [email protected] 6.6034.033 Metrodata IC Net 2.3 SR5 8.102.0043 Software Manua

Pagina 2 - All Rights Reserved

Table of contents IC Net 2.3 VIII 11.3 Index...363

Pagina 3 - Table of contents

6 Devices IC Net 2.3 90 6.5.4 Hardware settings Selection of the Hardware menu item of the 761 icon using the right mouse button opens the hardware

Pagina 4

6.5 761 Compact IC IC Net 2.3 91Cell constant Cell constant of the conductivity cell for cor-rect display of the absolute conductivity. Enter the

Pagina 5

6 Devices IC Net 2.3 92 Pump stop (P max/min) Pump stopped because pressure limits are exceeded. Hardware error Hardware error detected at the 761

Pagina 6

6.5 761 Compact IC IC Net 2.3 93Event Events for alarm stop: Pump stop Pump stopped because pressure limits are exceeded. Hardware error Hardw

Pagina 7

6 Devices IC Net 2.3 94 6.6 817 Bioscan 6.6.1 817 Bioscan features The Metrohm 817 Bioscan is a pulsed amperometric detector for use with a modular

Pagina 8

6.6 817 Bioscan IC Net 2.3 95 Mode Selection of detector mode: DC In DC mode, a constant potential is applied to the working electrode, the anal

Pagina 9

6 Devices IC Net 2.3 96 Valve Manual switching of the injection valve at the 812 Valve Unit: <Fill> Switch injection valve to "Fill&quo

Pagina 10 - IC Net 2.3

6.6 817 Bioscan IC Net 2.3 97Method data Cell potential Electric potential applied to the amperomet-ric detector cell in V Range: -2 ... +2 V. Ov

Pagina 11 - 1 Introduction

6 Devices IC Net 2.3 98 E1, E2, E3 Working potentials. Range: -2 ... +2 V. t1, t2, t3 Time duration to apply the working poten-tial. Range: 0.01 .

Pagina 12

6.6 817 Bioscan IC Net 2.3 99Actual E1, E2, E3 Display of current working potentials in V. t1, t2, t3 Display of current time durations for work

Pagina 13

1.1 Purpose of program IC Net 2.3 1 1 Introduction 1.1 Purpose of program «IC Net 2.3» is the name of the data acquisition and control software f

Pagina 14 - 1.4 Notation and pictograms

6 Devices IC Net 2.3 100 Range Selection of measuring range: pA 10 ... 5000 pA. nA 0.1 ... 50 nA. µA 0.01 ... 5 µA. Full scale The full

Pagina 15 - 1.5 Installation

6.6 817 Bioscan IC Net 2.3 101<Set> Send current parameters immediately to the 817 Bioscan. Parameters are not stored in the system file (*

Pagina 16

6 Devices IC Net 2.3 102 6.7 871 Advanced Bioscan 6.7.1 871 Advanced Bioscan features The Metrohm 871 Advanced Bioscan is a pulsed amperometric de

Pagina 17

6.7 871 Advanced Bioscan IC Net 2.3 103opened by selecting this menu option with the right mouse button or by double-clicking the 871 icon in the

Pagina 18

6 Devices IC Net 2.3 104 Scan The Scan mode allows potential/current curves to be recorded. Cell Manual control of amperometric detector cell: <

Pagina 19

6.7 871 Advanced Bioscan IC Net 2.3 105 Mode Selection of detector mode: DC In DC mode, a constant potential is applied to the working electrode

Pagina 20 - 2.2 Glossary

6 Devices IC Net 2.3 106 Actual Display of actual values. Mode Display of current detector mode. Oven Display of current ovent state. Control Dis

Pagina 21

6.7 871 Advanced Bioscan IC Net 2.3 107Startup values E Electric potential applied to the amperomet-ric detector cell in V Range: -2 ... +2 V. Ra

Pagina 22 - 2.4 Main window elements

6 Devices IC Net 2.3 108 tsample Measuring time at the end of working po-tential E1. Selection: 20 … (t1 - 60) ms (in steps of 20) Range Selectio

Pagina 23 - IC NET main window:

6.7 871 Advanced Bioscan IC Net 2.3 109Offset Display of current offset. Data rate Display of current data rate. Polarity Display of current po

Pagina 24 - 2.6 Overview of file types

1 Introduction IC Net 2.3 2 1.2 What’s new in «IC Net 2.3 SR5»? New Operating system With Service Release 5 operation of IC Net 2.3 under Windows

Pagina 25 - 2.7 Context-sensitive menus

6 Devices IC Net 2.3 110 E1 Start potential for sweep. Range: -2 ... +2 V. E2 End potential for sweep. Range: -2 ... +2 V. Rate Definition of sca

Pagina 26 - 2.9 Help

6.7 871 Advanced Bioscan IC Net 2.3 111Scan shell Clear display. Closes the scan file without saving. Load voltammogram. To open a saved vo

Pagina 27 - 3 Software settings

6 Devices IC Net 2.3 112 Stop voltammogram acquisition. Open Help file. Links The Links tab of the 871 Advanced Bioscan is used for COM port se

Pagina 28

6.8 733 IC Separation Center IC Net 2.3 1136.8 733 IC Separation Center 6.8.1 733 IC Separation Center features The Metrohm 733 IC Separation Ce

Pagina 29

6 Devices IC Net 2.3 114 with the 733 IC Separation Center, the 732 IC Detector and the 762 IC Interface can be found in the 762 Instructions for Us

Pagina 30

6.8 733 IC Separation Center IC Net 2.3 115 733 IC Separation Center connected to 732 733.0130 Separation Center connected to 762 Configuratio

Pagina 31 - 3.3 Audit Trail

6 Devices IC Net 2.3 116 Entry field Set automatic switching of the suppressor module after each "fill" or "inject" of valve A.

Pagina 32

6.8 733 IC Separation Center IC Net 2.3 117<Add> Add new program instruction. <Delete> Delete selected program instruction. <Veri

Pagina 33 - 3.4 Global settings

6 Devices IC Net 2.3 118 6.9 820 IC Separation Center 6.9.1 820 IC Separation Center features The Metrohm 820 IC Separation Center is a thermally

Pagina 34

6.9 820 IC Separation Center IC Net 2.3 119stalled Separation Center, if two detectors are connected to a 820 IC Separation Center with two inject

Pagina 35

1.3 Information about the software manual IC Net 2.3 3 1.3 Information about the software manual This 8.102.0001 Software Manual provides a compr

Pagina 36 - Ignore last data directory

6 Devices IC Net 2.3 120 6.9.3 820 IC Separation Center window 820 icon / Open The 820 IC Separation Center window for parameter settings is opened

Pagina 37 - 4 Systems

6.9 820 IC Separation Center IC Net 2.3 121Valve A fill/inject Display of current valve position.* <FILL> Switch valve A to "fill&q

Pagina 38

6 Devices IC Net 2.3 122 First column Time at which program instruction is ap-plied. Entry range: 0.0 ... 999.9 min If no time is entered, the pro

Pagina 39

6.9 820 IC Separation Center IC Net 2.3 123 Thermostat The Thermostat tab of the 820 IC Separation Center or 820.02X0 IC Separation Center windo

Pagina 40 - 4.2 System file handling

6 Devices IC Net 2.3 124 Start with startup hardware Start the Thermostat with the startup of the system hardware. Enabled The temperature of the c

Pagina 41 - 4.3 System functions

6.10 709 IC Pump IC Net 2.3 1256.10 709 IC Pump 6.10.1 709 IC Pump features The Metrohm 709 IC Pump is a serial dual piston high pressure pump f

Pagina 42

6 Devices IC Net 2.3 126 6.10.3 709 IC Pump window 709 icon / Open The 709 IC Pump window for parameter settings and manual con-trol is opened by s

Pagina 43

6.10 709 IC Pump IC Net 2.3 127value during operation, it is switched off immediately. At the same time, the Pressure field color in the 709 setti

Pagina 44

6 Devices IC Net 2.3 128 Correction factor Flow correction factor to adjust the dis-played flow to the actual flow measured with the aid of a measu

Pagina 45 - 4.4 System settings

6.11 818 IC Pump IC Net 2.3 1296.11 818 IC Pump 6.11.1 818 IC Pump features The Metrohm 818 IC Pump is a metal-free serial dual piston high pres

Pagina 46

1 Introduction IC Net 2.3 4 The 8.732.2013 IC Applications Collection contains all the Application Notes on the subject of ion chromatography and ca

Pagina 47 - 4.5 System state window

6 Devices IC Net 2.3 130 6.11.3 818 IC Pump window 818 icon / Open The 818 IC Pump window for parameter settings and manual con-trol is opened by s

Pagina 48

6.11 818 IC Pump IC Net 2.3 131used. If the pump exceeds the preset limit value during operation, it is switched off immediately. At the same time

Pagina 49

6 Devices IC Net 2.3 132 First column Time at which program instruction is ap-plied. Entry range: 0.0 ... 999.9 min If no time is entered, the pro

Pagina 50 - 5 Interfaces

6.11 818 IC Pump IC Net 2.3 133Pmax 0.0 ... 30.0 MPa Set maximum pressure limit for the high pressure pump to the desired value. Pmin 0.0 ... 3

Pagina 51

6 Devices IC Net 2.3 134 6.12 Metrohm 709 Solvent Delivery Unit 6.12.1 Metrohm 709 SDU features A Metrohm 709 SDU (Solvent Delivery Unit) system

Pagina 52 - 5.2 762 IC Interface

6.12 Metrohm 709 Solvent Delivery Unit IC Net 2.3 135Manual The Manual tab of the Pump Control Unit window is only available for a connected syste

Pagina 53

6 Devices IC Net 2.3 136 Stop after Stop time for all pump drives. Entry range: 0 ... 99999 min <Set> Send current parameters immediately t

Pagina 54

6.12 Metrohm 709 Solvent Delivery Unit IC Net 2.3 137Program table Number Row number in the program table (read-only). Time Time at which the fl

Pagina 55

6 Devices IC Net 2.3 138 Interfaces The Interfaces tab of the Pump Control Unit window is only avail-able for a connected system. Number of eluent

Pagina 56

6.13 Metrohm 818 Solvent Delivery Unit IC Net 2.3 1396.13 Metrohm 818 Solvent Delivery Unit 6.13.1 Metrohm 818 SDU features A Metrohm 818 SDU (S

Pagina 57

1.5 Installation IC Net 2.3 5 1.5 Installation 1.5.1 Hardware requirements for the PC Computer Pentium III with 450 MHz or higher Operating syst

Pagina 58

6 Devices IC Net 2.3 140 The Manual tab of the Pump Control Unit window is only available for a connected system. Pressure Live display of curren

Pagina 59 - 5.3 830 IC Interface

6.13 Metrohm 818 Solvent Delivery Unit IC Net 2.3 141Stop after Stop time for all pump drives. Entry range: 0 ... 99999 min <Set> Send cu

Pagina 60

6 Devices IC Net 2.3 142 Flow Overall flow of pump control system in mL/min. Entry range: 0.01 ... 20.00 mL/min (depending on number of pumps) %B

Pagina 61

6.13 Metrohm 818 Solvent Delivery Unit IC Net 2.3 143Pumps List of used pumps for gradient mixing. Clicking on <A>, <B>, <C>, o

Pagina 62

6 Devices IC Net 2.3 144 6.13 752 Pump Unit 6.13.1 752 Pump Unit features The Metrohm 752 Pump Unit is a 2-channel peristaltic pump which is norma

Pagina 63

6.13 752 Pump Unit IC Net 2.3 145Manual The Manual tab of the 752 Pump Unit window for manual operation is only available for a connected system.

Pagina 64

6 Devices IC Net 2.3 146 ENABLED Enable program start (a disabled program is not started). <Add> Add new program instruction. <Delete>

Pagina 65

6.14 753 Suppressor Module IC Net 2.3 1476.14 753 Suppressor Module 6.14.1 753 Suppressor Module features The Metrohm 753 Suppressor Module cons

Pagina 66

6 Devices IC Net 2.3 148 Manual The Manual tab of the 753 Suppressor Module window for manual operation is only available for a connected system. P

Pagina 67 - 1 Line on, active (0 V)

6.14 753 Suppressor Module IC Net 2.3 149First column Time at which program instruction is ap-plied. Entry range: 0.0 ... 999.9 min If no time is

Pagina 68 - 5.4 771 Compact Interface

1 Introduction IC Net 2.3 6 To change from an IC Net installation (without 21 CFR part 11 features) to an IC Net Compliant (with 21 CFR part 11 fe

Pagina 69

6 Devices IC Net 2.3 150 6.15 754 Dialysis Unit 6.15.1 754 Dialysis Unit features The Metrohm 754 Dialysis Unit is a unit for on-line sample dialy

Pagina 70 - 5.5 Metrohm PC Board

6.15 754 Dialysis Unit IC Net 2.3 151Manual The Manual tab of the 754 Dialysis Unit window for manual opera-tion is only available for a connected

Pagina 71 - 5.6 Global timer

6 Devices IC Net 2.3 152 Third column Parameter for program instruction (see be-low). ENABLED Enable program start (a disabled program is not star

Pagina 72

6.16 793 Sample Prep Module IC Net 2.3 1536.16 793 Sample Prep Module 6.16.1 793 Sample Prep Module features The Metrohm 793 Sample Prep Module

Pagina 73 - 6 Devices

6 Devices IC Net 2.3 154 Manual The Manual tab of the 793 Sample Prep Module window for manual operation is only available for a connected system.

Pagina 74

6.16 793 Sample Prep Module IC Net 2.3 155First column Time at which program instruction is ap-plied. Entry range: 0.0 ... 999.9 min If no time i

Pagina 75 - 6.2 Watch window

6 Devices IC Net 2.3 156 6.17 812 Valve Unit 6.17.1 812 Valve Unit features The Metrohm 812 Valve Unit is a control unit with one or two built-in

Pagina 76

6.17 812 Valve Unit IC Net 2.3 1576.17.3 812 Valve Unit window 812 icon / Open The 812 Valve Unit window for parameter settings is opened by se-l

Pagina 77

6 Devices IC Net 2.3 158 First column Time at which program instruction is ap-plied. Entry range: 0.0 ... 999.9 min If no time is entered, the prog

Pagina 78 - 6.3 732 IC Detector

6.18 816 IC Eluent Selector IC Net 2.3 1596.18 816 IC Eluent Selector 6.18.1 816 IC Eluent Selector features The Metrohm 816 IC Eluent Selector

Pagina 79

1.5 Installation IC Net 2.3 7 Reports Folder for storage of report files (*) and graphic files (*.wmf). Services Folder for user administration.

Pagina 80

6 Devices IC Net 2.3 160 Manual The Manual tab of the 816 IC Eluent Selector window for manual operation is only available for a connected system.

Pagina 81

6.18 816 IC Eluent Selector IC Net 2.3 161 First column Time at which program instruction is ap-plied. Entry range: 0.0 ... 999.9 min If no time

Pagina 82

6 Devices IC Net 2.3 162 6.19 828 IC Dual Suppressor 6.19.1 828 IC Dual Suppressor features The 828 IC Dual Suppressor is a continuous, regenerati

Pagina 83

6.19 828 IC Dual Suppressor IC Net 2.3 163Manual The Manual tab of the 828 IC Dual Suppressor window for manual operation is only available for a

Pagina 84

6 Devices IC Net 2.3 164 First column Time at which program instruction is ap-plied. Entry range: 0.0 ... 999.9 min If no time is entered, the pro

Pagina 85 - 6.4 819 IC Detector

6.20 833 IC Liquid Handling Unit IC Net 2.3 1656.20 833 IC Liquid Handling Unit 6.20.1 833 IC Liquid Handling Unit features The Metrohm 833 Liqu

Pagina 86

6 Devices IC Net 2.3 166 Open Open the 833 Liquid Handling Unit window for parame-ter settings (this window can also be opened by dou-ble-clicking t

Pagina 87

6.20 833 IC Liquid Handling Unit IC Net 2.3 167Suppressor (available only for 833 IC Suppressor Unit) <Step> Rotate suppressor to the next

Pagina 88

6 Devices IC Net 2.3 168 If no time is entered, the program instruc-tion is applied together with the last instruc-tion with time entry. Second colu

Pagina 89

6.21 837 IC Degasser IC Net 2.3 1696.21 837 IC Degasser 6.21.1 837 IC Degasser features The 837 IC Degasser is an instrument for continuously de

Pagina 90

1 Introduction IC Net 2.3 8 8. Copy the Autodatabase backup files of the folders Filters, Re-port templates, and User database into the folders ..\

Pagina 91

6 Devices IC Net 2.3 170 6.21.3 837 IC Degasser window 828 icon / Open The 837 IC Degasser window for parameter settings is opened by selecting th

Pagina 92 - 6.5 761 Compact IC

6.21 837 IC Degasser IC Net 2.3 171Time program On the Program tab of the 837 IC Degasser window a user-defined time program can be entered. This

Pagina 93

6 Devices IC Net 2.3 172 List of program instructions The following program instructions can be added to the time pro-gram on the Program page: Inst

Pagina 94

6.22 750 Autosampler IC Net 2.3 1736.22 750 Autosampler 6.22.1 750 Autosampler features The Metrohm 750 Autosampler is a sampler for automatic sa

Pagina 95

6 Devices IC Net 2.3 174 6.22.3 750 Autosampler window 750 icon / Open The 750 Autosampler window for parameter settings is opened by selecting thi

Pagina 96

6.22 750 Autosampler IC Net 2.3 175Loop size is Size of the sample loop at the 733 IC Sepa-ration Center. If the loop size is > 20 µL, only 3

Pagina 97

6 Devices IC Net 2.3 176 Manual injection Sample vial number Number of sample vial to be used for man-ual injection. <Select> Selection of th

Pagina 98

6.23 766 IC Sample Processor IC Net 2.3 1776.23 766 IC Sample Processor 6.23.1 766 IC Sample Processor features The Metrohm 766 IC Sample Proces

Pagina 99

6 Devices IC Net 2.3 178 If the system is connected and the 766 icon is clicked with the right mouse button, the following menu appears: Open Open t

Pagina 100

6.23 766 IC Sample Processor IC Net 2.3 179<Stop> Stop moving the rack to place the specified vessel at the needle. Position Selection of

Pagina 101

2.1 Starting/closing the program IC Net 2.3 9 2 Fundamentals of the operation 2.1 Starting/closing the program 2.1.1 Start the «IC Net» program

Pagina 102

6 Devices IC Net 2.3 180 Status display The current status of the 766 IC Sample Processor can be dis-played on the Current state subtab of the Manua

Pagina 103

6.23 766 IC Sample Processor IC Net 2.3 181 Num Row number (read-only). Task Program instruction (see below). Parameter Parameter for program i

Pagina 104 - 6.6 817 Bioscan

6 Devices IC Net 2.3 182 Ctrl "control command" Set the specified remote output signal. In addition to the normal control commands (detai

Pagina 105

6.23 766 IC Sample Processor IC Net 2.3 183Vial positions Rack type Current rack type (read-only). Parameter Name for special vial position. Pa

Pagina 106

6 Devices IC Net 2.3 184 Control Parameter Name for special control command. Pattern Remote line settings for special control command. <Add>

Pagina 107 - Display of current offset

6.23 766 IC Sample Processor IC Net 2.3 185Scan Parameter Name for special scan command. Pattern Remote line settings for special scan com-mand

Pagina 108

6 Devices IC Net 2.3 186 6.24 788 IC Filtration Sample Processor 6.24.1 788 IC Filtration Sample Processor features The Metrohm 788 IC Filtration

Pagina 109

6.24 788 IC Filtration Sample Processor IC Net 2.3 187If the system is connected and the 788 icon is clicked with the right mouse button, the foll

Pagina 110

6 Devices IC Net 2.3 188 <Stop> Stop moving the rack to place the specified vessel at the needle. Position Selection of the vessel to be pla

Pagina 111

6.24 788 IC Filtration Sample Processor IC Net 2.3 189Status display The current status of the 788 IC Filtration Sample Processor can be displayed

Pagina 112 - 6.7 871 Advanced Bioscan

All Rights Reserved. Printed in Switzerland by Metrohm AG, CH-9101 Herisau 2008

Pagina 113

2 Fundamentals of the operation IC Net 2.3 10 2.2 Glossary Batch reprocessing Batch reprocessing is understood to be the subsequent reproc-essing of

Pagina 114

6 Devices IC Net 2.3 190 Num Row number (read-only). Task Program instruction (see below). Parameter Parameter for program instruction (see be-l

Pagina 115

6.24 788 IC Filtration Sample Processor IC Net 2.3 191Ctrl "control command" Set the specified remote output signal. In addition to the

Pagina 116

6 Devices IC Net 2.3 192 Vial positions Rack type Current rack type (read-only). Parameter Name for special vial position. Pattern Number of spe

Pagina 117

6.24 788 IC Filtration Sample Processor IC Net 2.3 193Control Parameter Name for special control command. Pattern Remote line settings for spec

Pagina 118

6 Devices IC Net 2.3 194 Scan Parameter Name for special scan command. Pattern Remote line settings for special scan com-mand. <Add> Add a

Pagina 119

6.25 Metrohm Spark Triathlon Autosampler IC Net 2.3 1956.25 Metrohm Spark Triathlon Autosampler 6.25.1 Triathlon Autosampler features The Metroh

Pagina 120

6 Devices IC Net 2.3 196 6.25.3 Triathlon Autosampler window Triathlon icon / Open The Triathlon Autosampler window for parameter settings and man

Pagina 121

6.25 Metrohm Spark Triathlon Autosampler IC Net 2.3 197Rotate Sample Tray Rotate sample tray to access the tray seg-ments. The segments can be a

Pagina 122

6 Devices IC Net 2.3 198 Advanced The manual control of the advanced features can be done on the Advanced subtab of the Manual tab. Code out Combo

Pagina 123 - SYSTEM window of

6.25 Metrohm Spark Triathlon Autosampler IC Net 2.3 199Tray cooling The tray cooling is normally not included with the Metrohm Spark Triathlon Au

Pagina 124

2.2 Glossary IC Net 2.3 11fluenced by the integration parameters and the optionally pro-grammable integration events which are defined in the meth

Pagina 125

6 Devices IC Net 2.3 200 For detailed information about the Triathlon Autosampler see Spark Triathlon user manual.

Pagina 126

6.25 Metrohm Spark Triathlon Autosampler IC Net 2.3 201System settings This tab contains the configuration of the Triathlon Autosampler which is o

Pagina 127

6 Devices IC Net 2.3 202 Syringe volume, µL Volume of the syringe. Entry range: 100, 250, 500, 1000, 10000. Use the 10000 µL syringe only if the

Pagina 128 - 819 icons and two 820 icons

6.25 Metrohm Spark Triathlon Autosampler IC Net 2.3 203<Set> Set the specified values. For detailed information about the Triathlon Autosam

Pagina 129

6 Devices IC Net 2.3 204 <Select> Open the Select segment window to select a tray segment. <Set> Apply the segments selected under Meth

Pagina 130

6.25 Metrohm Spark Triathlon Autosampler IC Net 2.3 205Methods programming This tab contains the configuration of pre-defined Triathlon meth-ods a

Pagina 131

6 Devices IC Net 2.3 206 User program Enable the execution of the free program-mable User program, the Triathlon methods are disabled. User specifi

Pagina 132

6.25 Metrohm Spark Triathlon Autosampler IC Net 2.3 207Triathlon methods The Methods programming/Triathlon methods tab of the Metrohm Spark Triath

Pagina 133

6 Devices IC Net 2.3 208 Entry range: 0…999 µL, default: 30 (with air segment) The minimum flush volume depends on the volume of needle and needle

Pagina 134

6.25 Metrohm Spark Triathlon Autosampler IC Net 2.3 209 The Methods programming/Triathlon methods/Injection/Pick-up meode’s options subtab of the

Pagina 135 - 6.10 709 IC Pump

2 Fundamentals of the operation IC Net 2.3 12 2.3 Overview of program windows «IC Net 2.3» consists of different windows whose functionality is lin

Pagina 136

6 Devices IC Net 2.3 210 Last Transport Vial Combo box to specify the last transport vial position. Only active if Transport Vial = speci-fied samp

Pagina 137

6.25 Metrohm Spark Triathlon Autosampler IC Net 2.3 211Method values Combo box to select the parameters of the Wash method. Method Parameter Wa

Pagina 138

6 Devices IC Net 2.3 212 tor P5 AUXILLIARIES. The time program for each control line is arranged on a sepa-rate sheet. AUX-1,…AUX-4 Sheet for the t

Pagina 139 - 6.11 818 IC Pump

6.25 Metrohm Spark Triathlon Autosampler IC Net 2.3 213IO-output codes Subtab for the settings to control the hexa-decimal output at conncetor P3

Pagina 140

6 Devices IC Net 2.3 214 End time Subtab for the end time settings of the timebase program. Current Displays the current End time of the Triath-lo

Pagina 141

6.25 Metrohm Spark Triathlon Autosampler IC Net 2.3 215<Set> Apply values selected under Method values to the End time, the set values are

Pagina 142

6 Devices IC Net 2.3 216 Current Displays the current action sequence of the Triathlon Autosampler. The method’s action lot Programmable action se

Pagina 143

6.25 Metrohm Spark Triathlon Autosampler IC Net 2.3 217 Programmable actions:common specificEnd None Only for Mix method Aspirate from Sample Repe

Pagina 144

6 Devices IC Net 2.3 218 Vials for Userprog and Mix Method The Methods programming/Triathlon methods/Vials for Userprog and Mix method tab of the Tr

Pagina 145

6.25 Metrohm Spark Triathlon Autosampler IC Net 2.3 219 Position number of the transport vial, is successively increased by one as a sam-ple queu

Pagina 146

2.5 Icons of the main window IC Net 2.3 132.5 Icons of the main window The following icons are displayed in the IC NET main window: Open chromat

Pagina 147

6 Devices IC Net 2.3 220 6.26 838 Advanced Sample Processor 6.26.1 838 Advanced Sample Processor features The Metrohm 838 Advanced Sample Processo

Pagina 148

6.26 838 Advanced Sample Processor IC Net 2.3 221Unlink Delete the 838 icon from the system. 6.26.3 838 Advanced Sample Processor window 838 ico

Pagina 149

6 Devices IC Net 2.3 222 Selection: sample, sample+1, ext.#, next, prev, +rotate, -rotate, +swing, -swing, 1...151 sample: move to the position set

Pagina 150

6.26 838 Advanced Sample Processor IC Net 2.3 223Number Selection of the rack position. Selection: 1...n For detailed information about the 838 A

Pagina 151

6 Devices IC Net 2.3 224 empty: Tubing system and dosino cylinder are emptied. endVol: Dosino cylinder contents are ejected. compen.: Cancels the

Pagina 152

6.26 838 Advanced Sample Processor IC Net 2.3 225Status display The current status of the 838 Advanced Sample Processor can be displayed on the Cu

Pagina 153

6 Devices IC Net 2.3 226 For detailed information about the 838 Advanced Sample Proc-essor see 838 Instructions for Use. Time program On the Progra

Pagina 154 - 6.13 752 Pump Unit

6.26 838 Advanced Sample Processor IC Net 2.3 227List of program instructions The following program instructions can be added to the time pro-gram

Pagina 155

6 Devices IC Net 2.3 228 lution (D) in the sample queue: Concentr. vol. = MV/D Dilution vol. Dose Dilution vol. (DV) = MV * (D-1)/D Reagent vol. D

Pagina 156

6.26 838 Advanced Sample Processor IC Net 2.3 229Configuration The Configuration tab of the 838 Advanced Sample Processor win-dow contains parame

Pagina 157 - 6.14 753 Suppressor Module

2 Fundamentals of the operation IC Net 2.3 14 2.6 Overview of file types The following file types are produced by the «IC Net» software: *.adb Au

Pagina 158

6 Devices IC Net 2.3 230 Click button <Add> to open the ADD NEW RACK DATA window and add a new rack. Settings Rotate angle Definition of t

Pagina 159

6.26 838 Advanced Sample Processor IC Net 2.3 231Vial positions Special positions Definition of special positions. Vial number Number of the sp

Pagina 160 - 6.15 754 Dialysis Unit

6 Devices IC Net 2.3 232 Needle positions Needle positions Height, mm Lift position of special needle position in mm. <Add> Add a new special

Pagina 161

6.26 838 Advanced Sample Processor IC Net 2.3 233Control Parameter Name for special control command. Pattern Remote line settings for special c

Pagina 162

6 Devices IC Net 2.3 234 Scan Parameter Name for special scan command. Pattern Remote line settings for special scan com-mand. <Add> Add a

Pagina 163 - 6.16 793 Sample Prep Module

6.26 838 Advanced Sample Processor IC Net 2.3 235Periphery configuration To edit the configuration of dosinos, stirrers and membrane pumps, click

Pagina 164

6 Devices IC Net 2.3 236 Alias Define the alias name which appears in the rest of the driver. Default: Dosino MSB1…3 Dos port Select Dosino port f

Pagina 165

6.26 838 Advanced Sample Processor IC Net 2.3 237Alias Define the alias name which appears in the rest of the driver. Default: Stirrer MSB1…3, T

Pagina 166 - 6.17 812 Valve Unit

6 Devices IC Net 2.3 238 Links The Links tab of the 838 Advanced Sample Processor is used for COM port selection and settings (details see section

Pagina 167

6.27 861 Advanced Compact IC IC Net 2.3 239The 861 icon is available in the system window if a 861 Advanced Compact IC has been installed with the

Pagina 168

2.7 Context-sensitive menus IC Net 2.3 15*.rtt Report file (ASCII file) Contains a report template. The *.rtt file is stored in the program folde

Pagina 169 - 6.18 816 IC Eluent Selector

6 Devices IC Net 2.3 240 Conductivity, µS/cm Live display of measured conductivity. Pressure, MPa Live display of measured pressure. The color se

Pagina 170

6.27 861 Advanced Compact IC IC Net 2.3 241Peristaltic Peristaltic pump. (inactive with version 2.861.0010) <On> Start peristaltic pump. &l

Pagina 171

6 Devices IC Net 2.3 242 The Program tab contains the two following subpages: Program Main time program with all program steps. Remote configurati

Pagina 172 - 6.19 828 IC Dual Suppressor

6.27 861 Advanced Compact IC IC Net 2.3 243Instruction Parameter entry Meaning Valve Inject, Fill Switch injection valve to "INJECT" or

Pagina 173

6 Devices IC Net 2.3 244 Name (1st column) User-definable name of the remote com-mand (e.g. Start_838). Remote command (2nd column) Setting the re

Pagina 174

6.27 861 Advanced Compact IC IC Net 2.3 245 861 unit version Selection of instrument version: Suppressor Suppressor module: Autostep with Automa

Pagina 175

6 Devices IC Net 2.3 246 The Hardware tab of the hardware settings window defines ge-neral parameters which are set automatically at power on of the

Pagina 176

6.27 861 Advanced Compact IC IC Net 2.3 247For a precise determination of the cell con-stant, pump a calibration solution of known conductivity th

Pagina 177

6 Devices IC Net 2.3 248 1 (line on, active, 0 V) p (pulse output) * (leave line in current status) PowerOn values Display of the power o

Pagina 178

6.28 844 UV/VIS Compact IC IC Net 2.3 2496.28 844 UV/VIS Compact IC 6.28.1 844 UV/VIS Compact IC features The Metrohm 844 UV/VIS Compact IC is a

Pagina 179 - 6.21 837 IC Degasser

2 Fundamentals of the operation IC Net 2.3 16 [ PageDown ] Shifts a chromatogram on 1/10 part of a screen downwards. [ Shift ] + [ up ] Increases

Pagina 180

6 Devices IC Net 2.3 250 Controls On the subtab Controls injection valve, peristaltic pump, high-pressure pump, column heating and remote lines can

Pagina 181

6.28 844 UV/VIS Compact IC IC Net 2.3 25125.0 MPa (250 bar). Pressure min, MPa Value for minimum switch-off pressure for high-pressure pump. Once

Pagina 182 - Degasser on, off

6 Devices IC Net 2.3 252 <Off> Turn VIS lamp off. <Zero> Records a new zero spectrum. The zero spectrum is a reference spectrum. It sh

Pagina 183 - 6.22 750 Autosampler

6.28 844 UV/VIS Compact IC IC Net 2.3 253Current state Channel 1 abs, mAU Live display of the absorption in channel 1. Channel 2 abs , mAU Live

Pagina 184

6 Devices IC Net 2.3 254 determination) or at the moment the sample is injected (Start with inject). The Program tab contains the two following sub

Pagina 185

6.28 844 UV/VIS Compact IC IC Net 2.3 255<Delete> Delete selected program instruction. <Verify> Test the time program (error message

Pagina 186

6 Devices IC Net 2.3 256 Remote configuration On the Remote configuration subtab user-defined remote com-mands can be defined, which can be inserted

Pagina 187

6.28 844 UV/VIS Compact IC IC Net 2.3 257 Startup with hardware The activated devices will be started with Startup hardware or at the start of a

Pagina 188

6 Devices IC Net 2.3 258 6.28.4 Hardware settings Selection of the Hardware menu item using the right mouse button opens the hardware settings windo

Pagina 189

6.28 844 UV/VIS Compact IC IC Net 2.3 259are stopped immediately, the running determination and the ac-tive sample queue are also stopped. Shutdo

Pagina 190

3.1 Fonts IC Net 2.3 173 Software settings 3.1 Fonts IC NET / Options / Fonts This option allows the selection of fonts used by the system. Font

Pagina 191

6 Devices IC Net 2.3 260 System settings On the tab System settings break times for the injection valve and flow rate corrections for the high-press

Pagina 192

6.28 844 UV/VIS Compact IC IC Net 2.3 261Detector settings On the tab Detector settings measuring settings are defined. Detector settings Settin

Pagina 193

6 Devices IC Net 2.3 262 Operating hours Operating hours of the lamps. UV lamp, h Display of the operating hours of the UV lamp. <Reset UV lam

Pagina 194

6.28 844 UV/VIS Compact IC IC Net 2.3 263ample for the wavelength set for measuring channel 1: aktIntzeroIntAbs11logλλ= where: zeroInt1λ The int

Pagina 195

6 Devices IC Net 2.3 264 <Save as> Opens the dialog for entering the path and file name. Saves the last recorded spectrum in the selected .cv

Pagina 196

6.29 Data recorder IC Net 2.3 2656.29.2 Select processing method and data source The Data recorder icon / Setup menu item opens the following win

Pagina 197

6 Devices IC Net 2.3 266 # ch1 and ch2 Data source channels of the 771 Compact Interface. 761 Compact IC # Cond Conductivity channel of the 761 Com

Pagina 198

6.30 System timer IC Net 2.3 2676.30 System timer 6.30.1 Timer icon The system timer is used for programming system tasks which are started autom

Pagina 199

7 Methods IC Net 2.3 268 7 Methods A method contains all information necessary for data acquisi-tion, integration, peak evaluation and quantificatio

Pagina 200

7.2 Passport IC Net 2.3 2697.2.1 General Tab General of the passport with general description of the method and determination. Ident User defin

Pagina 201

3 Software settings IC Net 2.3 18 User Window Possible actions Novice, Master New Password This window allows the user to change his Password. Adm

Pagina 202

7 Methods IC Net 2.3 270 To reset the Run number press [Ctrl] [F8]. The Analysis number window is opened where the run number for the next run can

Pagina 203

7.2 Passport IC Net 2.3 271Internal standard amount Concentration of the internal standard component for relative concentration calcu-lations. Dat

Pagina 204

7 Methods IC Net 2.3 272 7.2.4 Eluent Tab Eluent of passport with eluent information. Eluent A Description of eluent composition (max. 256 charac

Pagina 205

7.2 Passport IC Net 2.3 2737.2.5 Comment Tab Comment of passport for entry of free-text user comments into the chromatogram description. Use this

Pagina 206

7 Methods IC Net 2.3 274 7.3 Method setup IC NET / Method / Method setup This menu item opens the Method setup window which includes the most comm

Pagina 207

7.3 Method setup IC Net 2.3 275Start delay Time delay before starting data acquisition in min. Detector name User-defined detector name. Data ac

Pagina 208

7 Methods IC Net 2.3 276 bit smaller and wider after smoothing, but their area does not change. Raw data points Number of data points per chromato

Pagina 209

7.3 Method setup IC Net 2.3 277 Optimize your analysis first to get stable measuring conditions without a retention time drift. In any case it is

Pagina 210

7 Methods IC Net 2.3 278 7.3.5 Export The tab Export of Method setup window contains the export op-tions for chromatograms that are performed during

Pagina 211

7.3 Method setup IC Net 2.3 279to use this feature. Depending on the Excel version in use, settings of Excel have to be modified. For detailed in

Pagina 212

3.2 Security system IC Net 2.3 19<Modify user> Edit current user's settings (username, full name, access level, status), opens the Mod

Pagina 213

7 Methods IC Net 2.3 280 European Pharmacopoeia Formula setting according to European Pharmacopoeia. US Pharmacopoeia Formula setting according to

Pagina 214

7.4 Integration IC Net 2.3 2817.4 Integration 7.4.1 General information IC NET / Method / Integration This menu item opens the Integration param

Pagina 215

7 Methods IC Net 2.3 282 7.4.2 Setup Tab Setup of Integration parameters window with integration pa-rameters. The two most important of these parame

Pagina 216

7.4 Integration IC Net 2.3 283width of the narrowest peak on the chroma-togram (normally 2…10 s). Entry range: 0.1…480 s Broadening Broadening f

Pagina 217

7 Methods IC Net 2.3 284 <Suggest> This function sets up reasonable integration parameters in the following way: • Slope is set to a reasona

Pagina 218

7.4 Integration IC Net 2.3 285If an integration event is added or modified, the Edit integration event window is opened where the following parame

Pagina 219

7 Methods IC Net 2.3 286 Integration events for peak start/end Set peak start Force the beginning of a new peak. If a peak is available at the mome

Pagina 220

7.4 Integration IC Net 2.3 287Set horizontal baseline Set horizontal baseline for all peaks except the last one of adjacent peaks that are not sep

Pagina 221

7 Methods IC Net 2.3 288 Force horizontal Cancel horizontal base back base back Force horizontal base back Set horizontal baseline for a si

Pagina 222

7.5 Calibration and quantification IC Net 2.3 2897.5 Calibration and quantification 7.5.1 General information The aim of any chromatographic anal

Pagina 223

Table of contents IC Net 2.3 I Table of contents 1 Introduction ...1 1.1 Purpose of p

Pagina 224

3 Software settings IC Net 2.3 20 failed Login attempts exceeds the limit set in the Password Options. The user can be set Active again by the Admin

Pagina 225

7 Methods IC Net 2.3 290 7.5.2 Notations R Stands for response value, either area or height, depending on setting selected in the Calibration graph

Pagina 226

7.5 Calibration and quantification IC Net 2.3 2917.5.3 External standard calibration External standard calibration (absolute calibration) is a ke

Pagina 227

7 Methods IC Net 2.3 292 Number Row number (read-only). Peak Number of the peak in the chromatogram that corresponds to the given component. If

Pagina 228

7.5 Calibration and quantification IC Net 2.3 293min C (max C) Minimum (Maximum) concentration value for the component. Components whose concentr

Pagina 229

7 Methods IC Net 2.3 294 7.5.5 Peak identification IC NET / Method / Calibration / Identification Selecting this menu item or clicking on <Identi

Pagina 230

7.5 Calibration and quantification IC Net 2.3 295Worst case... Information on the component with the worst (largest) deviation of actual and ex-p

Pagina 231

7 Methods IC Net 2.3 296 Level 1...Level N Contains concentrations (or other chosen values) of components for corresponding Calibration level. <

Pagina 232

7.5 Calibration and quantification IC Net 2.3 2977.5.7 Calibration curve IC NET / Method / Calibration / Graphs Selecting this menu item or clicki

Pagina 233

7 Methods IC Net 2.3 298 Calibration points table This table contains the basic information used to construct the calibration curve: Level Number

Pagina 234

7.5 Calibration and quantification IC Net 2.3 299Base Base for calculations (Area or Height) in quantification and calibration procedures. Indic

Pagina 235

3.3 Audit Trail IC Net 2.3 21Automatic Logout After After the specified period of inactivity (in min-utes) the system is locked, to protect it aga

Pagina 236

7 Methods IC Net 2.3 300 amount of the injected components becomes equal to the declared value. This declared sum-mary amount is entered into the Co

Pagina 237

7.6 Report output IC Net 2.3 3017.6 Report output 7.6.1 Report options window IC NET / Method / Report options IC NET / Process / Make report

Pagina 238

7 Methods IC Net 2.3 302 7.6.2 Items to report This part of the Report options window contains a list of important report items that can be include

Pagina 239

7.6 Report output IC Net 2.3 303Chromatogram plot Plot of the chromatogram (only for Printer or File destination). Peak table Peak table with

Pagina 240

7 Methods IC Net 2.3 304 7.6.3 More items to report This part of the Report options window contains a list of rarely used report items that can be

Pagina 241

7.6 Report output IC Net 2.3 305Component table Component table. No Retention Window% RF Conc. Index Type Group Name 1 3.0

Pagina 242

7 Methods IC Net 2.3 306 Name Channel’s name. This text appears as Channel label in the chromatogram. Units User-defined units of detector respo

Pagina 243

7.6 Report output IC Net 2.3 3077.6.4 Report destination The following output devices can be selected simultaneously in any combination as targets

Pagina 244

7 Methods IC Net 2.3 308 As long as no universal component is de-fined, only peaks with non-zero concentra-tions are listed. The peak table includes

Pagina 245

7.6 Report output IC Net 2.3 309Concentration of std Concentration of the internal standard component for relative concentration calcu-lations. T

Pagina 246

3 Software settings IC Net 2.3 22 Audit Trail window The Audit Trail window displays the Audit Trail records of the IC Net or Chromatogram History i

Pagina 247

7 Methods IC Net 2.3 310 7.6.6 Template options For report output using a report template the following parame-ters can be modified: Template Selec

Pagina 248

7.6 Report output IC Net 2.3 3117.6.8 Report elements If Quantification method = Custom has been selected, the peak ta-ble can be customized on th

Pagina 249

7 Methods IC Net 2.3 312 where i and i+1 indexes refer to the neigh-boring peaks, and w0.607 stands for the peak width at 60.7 % of the peak height.

Pagina 250

7.6 Report output IC Net 2.3 313PGF = 1 for an ideal peak with gaussian shape. asymmetry Peak asymmetry As is calculated at 1/10 of the peak heig

Pagina 251

7 Methods IC Net 2.3 314 peak amount Absolute amount of substance which is contained in the sample loop volume. It is the product of the absolute c

Pagina 252

7.6 Report output IC Net 2.3 315The complete component type may look like: BBD : !R. group Number of the group for the component. spectral ratio

Pagina 253

8 Chromatograms IC Net 2.3 316 8 Chromatograms 8.1 Chromatogram window The CHROMATOGRAM window is used to show a running or re-corded chromatogram.

Pagina 254

8.2 Chromatogram file handling IC Net 2.3 317During active data acquisition the elapsed time, the duration of the chromatogram or program and the

Pagina 255

8 Chromatograms IC Net 2.3 318 Filename Filename (common wildcards as * and ? are possible). Selected Number of selected chromatograms and their

Pagina 256

8.2 Chromatogram file handling IC Net 2.3 3198.2.2 Save chromatogram IC NET / File / Save / Chromatogram Save the selected chromatogram in a chr

Pagina 257

3.4 Global settings IC Net 2.3 23Sort window This window offers a more refined possibility to arrange log data by combining the column titles of t

Pagina 258

8 Chromatograms IC Net 2.3 320 Represent raw data as Output of raw data: Default units (float) Y values are exported in default units for the chan

Pagina 259 - 6.28 844 UV/VIS Compact IC

8.3 Graphical representation IC Net 2.3 3218.3 Graphical representation 8.3.1 Appearance IC NET / View / Appearance This menu item opens the App

Pagina 260

8 Chromatograms IC Net 2.3 322 Response axis Y from Beginning of the window on Y axis. Y to End of the window on Y axis. Tic marks Plot axes la

Pagina 261

8.3 Graphical representation IC Net 2.3 323Component name Name of the component. Component name + Quantity Name and quantity of the component. C

Pagina 262

8 Chromatograms IC Net 2.3 324 Colors Tab Colors of the Appearance window with color settings parame-ters for chromatograms. <Choose> Choose

Pagina 263

8.3 Graphical representation IC Net 2.3 325By selecting this menu item or pressing [ Ctrl ] + [ Home ] the X axis is scaled so that the chromatogr

Pagina 264

8 Chromatograms IC Net 2.3 326 8.4 Peak editor 8.4.1 Switching on/off the peak editor IC NET / Process / Peak editor The peak editor mode is used

Pagina 265

8.4 Peak editor IC Net 2.3 327 IC NET / Peak / Select end point Move the cursor to the end of the nearest peak and select the peak. IC NET / Pe

Pagina 266

8 Chromatograms IC Net 2.3 328 8.5 Printing 8.5.1 Page layout for printing IC NET / File / Page layout The selection of this menu item or clicking

Pagina 267

8.5 Printing IC Net 2.3 329<Save as default> Save set page layout parameters as default values. 8.5.2 Printer settings IC NET / File / Pr

Pagina 268

3 Software settings IC Net 2.3 24 If method changed Global settings for saving of method files: Don’t ask to save The method file is not saved a

Pagina 269

8 Chromatograms IC Net 2.3 330 8.6 Miscellaneous functions 8.6.1 Reintegration IC NET / Process / Reintegrate This menu item opens the Integration

Pagina 270

8.6 Miscellaneous functions IC Net 2.3 331Meaning Meaning of the signature, can be selected from a pop-up list that is predefined in the Meaning

Pagina 271

8 Chromatograms IC Net 2.3 332 IC NET / Process / Electronic signature This window displays the signatures of the chromatogram in the active windo

Pagina 272

8.6 Miscellaneous functions IC Net 2.3 3338.6.5 Compare chromatogram IC NET / Process / More / Compare This menu item opens the Compare with win

Pagina 273

8 Chromatograms IC Net 2.3 334 File Close Close window and return to Compare with win-dow. Edit Copy Copy the data of the comparison to the clip-b

Pagina 274 - 6.29 Data recorder

8.6 Miscellaneous functions IC Net 2.3 335 8.6.7 Invert chromatogram IC NET / Process / More / Invert! This menu item inverts the response curve

Pagina 275

8 Chromatograms IC Net 2.3 336 <Cancel> Close the window and abandon changes. For details on the «Autodatabase 1.0» program, open this pro-g

Pagina 276

8.6 Miscellaneous functions IC Net 2.3 337Excel file The Excel file ExcelReport.xls is located in the folder ..\IC Net\ExcelReport\. Formatting, h

Pagina 277 - 6.30 System timer

8 Chromatograms IC Net 2.3 338 Make sure in «IC Net» that the item corresponding to the first column always contains information, otherwise a line

Pagina 278 - 7 Methods

8.6 Miscellaneous functions IC Net 2.3 339 [Items] Specifies columns, their headers and the order of the col-umns in the data worksheet. [Ions] S

Pagina 279

3.4 Global settings IC Net 2.3 25 Enable Always, if you use «IC Net» together with an Online IC. In This case Method/Method setup/Export/Send data

Pagina 280 - • Amount

8 Chromatograms IC Net 2.3 340 [Item] TableStart = A6, Specifies the cell in the data worksheet of the ExcelReport.xls where the header of the fir

Pagina 281

8.6 Miscellaneous functions IC Net 2.3 341Rel.conc. Relative (Internal Standard) concen-tration. Index Linear or logarithmic index. Type Marks

Pagina 282

8 Chromatograms IC Net 2.3 342 The names of the components for which data is to be exported to the Excel file have to be exactly the same (case se

Pagina 283

9.1 Sample queue file handling IC Net 2.3 3439 Sample queue A sample queue is a table containing sample-specific data which is used to facilitate

Pagina 284 - 7.3 Method setup

9 Sample queue IC Net 2.3 344 9.2 Sample queue control 9.2.1 Sample queue overview table IC NET / File / Open / Sample queue SYSTEM / System /

Pagina 285

9.2 Sample queue control IC Net 2.3 345Level Calibration level (see section 7.5) for the sample: Level 0 stays for normal analysis, levels 1 and

Pagina 286

9 Sample queue IC Net 2.3 346 Interrupt execution of the sample queue after the current determi-nation has been finished. A running sample queue can

Pagina 287

9.3 Sample queue editor IC Net 2.3 347 All system files of the sample queue must be stored in the same system subfolder. Ident User defined iden

Pagina 288

9 Sample queue IC Net 2.3 348 Cut the selected rows of the sample queue table and copy them into the clipboard. QUEUE EDITOR / Edit / Copy row(s)

Pagina 289

10.1 Batch reprocessing queue file handling IC Net 2.3 34910 Batch reprocessing Reprocessing is understood to be the subsequent reprocessing of a

Pagina 290

3 Software settings IC Net 2.3 26 Opening chromatograms Global settings for opening chroma-tograms: Ignore last data directory If this option is ena

Pagina 291 - 7.4 Integration

10 Batch reprocessing IC Net 2.3 350 10.1.3 Save batch reprocessing queue QUEUE EDITOR / File / Save Save the batch reprocessing queue in a batch

Pagina 292

10.2 Perform batch reprocessing IC Net 2.3 351 Reprocess calibration runs Reprocess all calibration chromatograms (calibration level > 0). Upda

Pagina 293

10 Batch reprocessing IC Net 2.3 352 switched on) by updating the concentration table. Default scheme Default setting for recalibration reprocess-i

Pagina 294

10.2 Perform batch reprocessing IC Net 2.3 353Internal standard amount entered in the batch reprocessing table. The recalculation is done automa

Pagina 295

10 Batch reprocessing IC Net 2.3 354 Send to Autodatabase Send chromatogram data to specified Autodatabase file. <Edit Autodatabase options>

Pagina 296

10.2 Perform batch reprocessing IC Net 2.3 355rel. concentration rel. concentration% peak amount Statistics Component Select the component which

Pagina 297

10 Batch reprocessing IC Net 2.3 356 Setting for the saving of the report file. Overwrite old report Write in new file (default) Ask before writi

Pagina 298

10.2 Perform batch reprocessing IC Net 2.3 35710.2.2 Merge chromatograms REPROCESS / <Merge> Combine all chromatograms of the batch reproces

Pagina 299 - Component table created

10 Batch reprocessing IC Net 2.3 358 10.3 Batch reprocessing queue editor 10.3.1 Open batch reprocessing queue editor window REPROCESS / <Edit

Pagina 300

10.3 Batch reprocessing queue editor IC Net 2.3 359Sample Info 1 First sample description. Will be placed into appropriate passport field. Sample

Pagina 301

4.1 System creation IC Net 2.3 274 Systems The combination of Metrohm and Bischoff devices connected to an interface is called a system. The syste

Pagina 302

11 Appendix IC Net 2.3 360 11 Appendix 11.1 Software license The use of the Software is subject to this License Agreement be-tween you and Metrohm A

Pagina 303

11.1 Software license IC Net 2.3 3615. Metrohm and its distributor disclaim all other warranties and liability. In particular, Metrohm and its

Pagina 304

11 Appendix IC Net 2.3 362 11.2 Declaration of conformity – Software validation The software "Metrodata IC Net 2.3" was developed in acco

Pagina 305

11.3 Index IC Net 2.3 36311.3 Index % %B 135,137,140,142 %C 135,137,140,142 %D 135,137,140,142 ( (multi)Channel ratio 306 < <Close> 264 &

Pagina 306

11 Anhang IC Net 2.3 364 window 43 766 IC Sample Processor configuration 182 features 177 icon 177 links 185 manual control 178 program 180 progr

Pagina 307

11.3 Index IC Net 2.3 365icon 165 links 168 manual settings 166 program 167 program instructions 168 window 166 833 IC Pump Unit features 165 ic

Pagina 308

11 Anhang IC Net 2.3 366 Add>47,48,55,56,72,74,80,81,86,88,117,122,132,137,142,146,149,152,155,158,161,164,168,171,181,183,184,185,190,192,193,

Pagina 309

11.3 Index IC Net 2.3 367function 290,299 general information 289 method 298 parameters 298 point 295,298 points table 298 recalibration 330 res

Pagina 310

11 Anhang IC Net 2.3 368 Coefficients 299 Colors 324 display settings 66 settings 324 system state window 66 watch window 66 Column 271,302 len

Pagina 311 - 7.6 Report output

11.3 Index IC Net 2.3 369Del> 62 Delay 282 Delay time 208 Delete>47,48,55,56,72,74,80,81,86,88,117,122,132,137,142,146,149,152,155,158,161,1

Pagina 312

4 Systems IC Net 2.3 28 After clicking on <Next>, the window for addition of interfaces to the toolbar appears. This step can be skipped if th

Pagina 313

11 Anhang IC Net 2.3 370 ENABLED47,55,72,79,86,116,122,132,146,149,152,155,158,168,242,254 END 48,55,87,243 End time 214 End> 197 Error message

Pagina 314

11.3 Index IC Net 2.3 371Group 292,355 Groups 309 H Hair 324 halfwidth 311 Handshake 45,52,59 Hardware 90,246,249 Hardware error 92,93,247,248,258

Pagina 315

11 Anhang IC Net 2.3 372 Insert> 216 Installation 5 Integration algorithm 281 definition 10 events 284 parameters 282 reintegration 330 Integra

Pagina 316

11.3 Index IC Net 2.3 373Merge> 354 Method definition 11,268 file handling 268 open 268 overwrite 24 path 269 processing method 29 save 24,26

Pagina 317

11 Anhang IC Net 2.3 374 On-line help 16 Open 249 Open all files> 350 Open example> 350 Opening chromatograms 26 Operating hours 262 Order

Pagina 318

11.3 Index IC Net 2.3 375Printer setup 329 Printing 328 Printing order 309 Printing range 329 Processing 276,316 Processing method 11,29,265,268

Pagina 319

11 Anhang IC Net 2.3 376 calibration curve 305 component table 305 directory for report output 310 elements 311 file name 310 file output 307

Pagina 320

11.3 Index IC Net 2.3 377Send data to IC Cap 278 Send data to Online 278 Send to Autodatabase 354 Send to Autodatabase> 335 Separator 310 Seria

Pagina 321

11 Anhang IC Net 2.3 378 file handling 30 folder 27,40,63 modify system window 35 new 27 open 30 overwrite 24 print system parameters 36 save 3

Pagina 322

11.3 Index IC Net 2.3 379Used 298 User 19,269 User administration 17 User comments 25 User List 18 User Name 17 User name> 330 User program 2

Pagina 323

4.1 System creation IC Net 2.3 29For every device to be controlled with «IC Net», open the group which contains this device, select the device and

Pagina 324

Table of contents IC Net 2.3 II 4.1.3 Add devices to an existing window... 30 4.2 System file handling ...

Pagina 325

4 Systems IC Net 2.3 30 4.1.3 Add devices to an existing window SYSTEM / Setup / New devices/ Install new device Add new devices to the open SY

Pagina 326 - 8 Chromatograms

4.3 System functions IC Net 2.3 31Save the current system file (*.smt) in the desired folder of the Systems directory. SYSTEM / System / Close Clo

Pagina 327

4 Systems IC Net 2.3 32 • Start Bischoff detectors. • Switch Metrohm Autosamplers and Bischoff Autosamplers to remote control. • Start recording

Pagina 328

4.3 System functions IC Net 2.3 33 If the Auto restart option is enabled or if a sample queue is active, a new determination is started immedi-ate

Pagina 329

4 Systems IC Net 2.3 34 chromatogram window and in the Chroma-togram open window. Calibration level Calibration level (0 = sample; 1…n = cali-brati

Pagina 330

4.4 System settings IC Net 2.3 354.4 System settings 4.4.1 Modify system window SYSTEM / Setup / Drag icons If this option is enabled, the device

Pagina 331 - 8.3 Graphical representation

4 Systems IC Net 2.3 36 Start with inject The time program of the loaded object or the data acquisition (RECORDER) is started at the moment the sam

Pagina 332

4.5 System state window IC Net 2.3 374.4.7 Show system parameters SYSTEM / Setup / Parameters / Startup hardware Display of the name of the metho

Pagina 333

4 Systems IC Net 2.3 38 Running Running program or chromatogram data acquisition. Running program (xxx min left) Running time program with time dis

Pagina 334

4.5 System state window IC Net 2.3 39E295 Storage of memory values failed (repeat last action; if error reappears, call Metrohm service). E296 Ins

Pagina 335

Table of contents IC Net 2.3 III 6 Devices ...63 6.1 Device installation ...

Pagina 336 - 8.4 Peak editor

5 Interfaces IC Net 2.3 40 5 Interfaces The «IC Net» program supports remote control for interfaces available from Metrohm and Bischoff. Interfaces

Pagina 337

5.1 Interface installation IC Net 2.3 41Metrohm Interfaces or Bischoff Interfaces groups. After selecting the serial port where the interface is c

Pagina 338 - 8.5 Printing

5 Interfaces IC Net 2.3 42 Click on <Install device> to open the Adding interfaces to your work-place window and proceed as described in secti

Pagina 339

5.2 762 IC Interface IC Net 2.3 435.2.3 762 icons in the system window The 762 System 1 and 762 System 2 icons are available in the SYS-TEM wi

Pagina 340 - 8.6 Miscellaneous functions

5 Interfaces IC Net 2.3 44 Measuring per second Number of data points measured per s. Entry range: 10, 20, 30, 50, 60 points/s Range Range for AD

Pagina 341

5.2 762 IC Interface IC Net 2.3 45Protected Protocol Data transfer with Protected Protocol be-tween 830 IC Interface and PC for enhanced data se

Pagina 342

5 Interfaces IC Net 2.3 46 Put on desktop Possibility for setting COM port parameters (details see on-line help). Change Possibility for changing

Pagina 343

5.2 762 IC Interface IC Net 2.3 47The Program tab contains the two following subpages: Program Main time program with all program steps. Events c

Pagina 344

5 Interfaces IC Net 2.3 48 Program END, RESET The END flag can be used to end a program, especially if the program time should be longer than the c

Pagina 345

5.3 830 IC Interface IC Net 2.3 49 Events setup The Events setup tab of the 762 System [#] window contains the startup values for the seven event

Pagina 346

Table of contents IC Net 2.3 IV 6.12.1 Metrohm 709 SDU features... 134 6.12.2 Metrohm 709 SDU icon ..

Pagina 347

5 Interfaces IC Net 2.3 50 5.3.2 830 icon on the toolbar The 830 icon is available on the toolbar if a 830 IC Interface has been installed with th

Pagina 348

5.3 830 IC Interface IC Net 2.3 51 Label Optional label to name the interface with maximum 8 characters. ROM version Instrument program version

Pagina 349

5 Interfaces IC Net 2.3 52 Protected Protocol Data transfer with Protected Protocol be-tween 830 IC Interface and PC for enhanced data security.

Pagina 350

5.3 830 IC Interface IC Net 2.3 53Links The Links tab of the 830 IC Interface is used for COM port selection and settings. Alias Name of the int

Pagina 351

5 Interfaces IC Net 2.3 54 5.3.5 830 Event output lines 830 System [#] icon / Open The 830 System [#] window with programs and settings for the sev

Pagina 352

5.3 830 IC Interface IC Net 2.3 55Program On the Program subpage, program steps including time, program instruction and parameter can be entered.

Pagina 353 - 9 Sample queue

5 Interfaces IC Net 2.3 56 Events configuration On the Events configuration subtab user-defined event commands can be defined, which can be inserte

Pagina 354 - 9.2 Sample queue control

5.3 830 IC Interface IC Net 2.3 57Events setup The Events setup tab of the 830 System [#] window contains the startup values for the seven event o

Pagina 355

5 Interfaces IC Net 2.3 58 5.4 771 Compact Interface 5.4.1 771 Compact Interface features The Metrohm 771 Compact Interface is a two channel Analog

Pagina 356 - 9.3 Sample queue editor

5.4 771 Compact Interface IC Net 2.3 59 Range Range for AD converter. Example: converts ± 2500 mV to 224 bits Entry range: 2500, 1250, 625, 312.

Pagina 357

Table of contents IC Net 2.3 V 6.23 766 IC Sample Processor... 177 6.23.1 766 IC Sample Pr

Pagina 358

5 Interfaces IC Net 2.3 60 <ASK> Read current parameters from 771 Compact Interface. <SET> Send current parameters to 771 Compact Inte

Pagina 359 - 10 Batch reprocessing

5.6 Global timer IC Net 2.3 615.6 Global timer 5.6.1 Timer icon The global timer is used for programming system tasks which are started automatic

Pagina 360

5 Interfaces IC Net 2.3 62 <Add subtask> Add a new subtask for the selected daily task. The time and program instruction for this sub-task ha

Pagina 361

6.1 Device installation IC Net 2.3 636 Devices The «IC Net» program supports remote control for Metrohm and Bischoff instruments. Each instrument

Pagina 362

6 Devices IC Net 2.3 64 After clicking on <Next>, the window for addition of interfaces to the toolbar appears (details see section 4.1.1). A

Pagina 363

6.2 Watch window IC Net 2.3 656.1.2 Install new devices in system window SYSTEM / Setup / New devices/ Install new device Add new devices to the

Pagina 364

6 Devices IC Net 2.3 66 6.2.2 Watch window settings Watch window icon / Open The WATCH WINDOW displays all live parameters of the interfaces and de

Pagina 365

6.2 Watch window IC Net 2.3 67measurement value is inside the normal range. Choose color / Background / Active Set display color for field backgro

Pagina 366

6 Devices IC Net 2.3 68 6.3 732 IC Detector 6.3.1 732 IC Detector features The Metrohm 732 IC Detector is a conductivity detector for ion chromato

Pagina 367

6.3 732 IC Detector IC Net 2.3 696.3.3 732 IC Detector window 732 icon / Open The 732 IC Detector window for parameter settings is opened by sele

Pagina 368

Table of contents IC Net 2.3 VI 7.3.5 Export... 278 7.3.6 Math...

Pagina 369

6 Devices IC Net 2.3 70 Range Display of measuring range in µS/cm. Full scale Display of full scale range in µS/cm. Cell const. Display of cell co

Pagina 370 - 11 Appendix

6.3 732 IC Detector IC Net 2.3 71Range Measuring range in µS/cm Selection: 100, 200, 500 µS/cm and 1, 2, 5, 10 mS/cm Full scale The full scale

Pagina 371

6 Devices IC Net 2.3 72 The Program tab contains the two following subpages: Program Main time program with all program steps. Remote configuratio

Pagina 372 - 11.3 Ordering designations

6.3 732 IC Detector IC Net 2.3 73List of program instructions The following program instructions can be added to the time pro-gram on the Program

Pagina 373 - 11.3 Index

6 Devices IC Net 2.3 74 Pmax 0.0 ... 25.0 MPa Set maximum pressure limit for the high-pressure pump con-nected to the 732 IC Detector to the desir

Pagina 374 - IC Net 2.3

6.4 819 IC Detector IC Net 2.3 756.4 819 IC Detector 6.4.1 819 IC Detector features The Metrohm 819 IC Detector is a conductivity detector for i

Pagina 375

6 Devices IC Net 2.3 76 Display The Display tab of the 819 IC Detector window is only available for a connected system. It displays the current meas

Pagina 376

6.4 819 IC Detector IC Net 2.3 77Damping Display of electronic damping of the analog output. Remote lines Display value of remote line settings

Pagina 377

6 Devices IC Net 2.3 78 Custom independent selection of polarity and tem-perature coeff. Polarity Polarity of the analog output signal: + positiv

Pagina 378

6.4 819 IC Detector IC Net 2.3 79 The Program tab contains the two following subpages: Program Main time program with all program steps. Remote c

Pagina 379

Table of contents IC Net 2.3 VII8.6 Miscellaneous functions... 330 8.6.1 Reintegration ..

Pagina 380

6 Devices IC Net 2.3 80 <Add> Add new program instruction. <Delete> Delete selected program instruction. <Verify> Test the time

Pagina 381

6.4 819 IC Detector IC Net 2.3 81Remote configuration On the Remote configuration subtab user-defined remote com-mands can be defined, which can b

Pagina 382

6 Devices IC Net 2.3 82 6.5 761 Compact IC 6.5.1 761 Compact IC features The Metrohm 761 Compact IC is a PC-controlled system for ion chromatograp

Pagina 383

6.5 761 Compact IC IC Net 2.3 83If the system is connected and the 761 icon is clicked with the right mouse button, the following menu appears: Op

Pagina 384

6 Devices IC Net 2.3 84 Conductivity, µS/cm Live display of measured conductivity. Pressure, MPa Live display of measured pressure. The color sett

Pagina 385

6.5 761 Compact IC IC Net 2.3 85even without connection to the PC. Entry range: 0.0 ... 25.0 MPa Full scale, µS/cm Startup value for full scale ra

Pagina 386

6 Devices IC Net 2.3 86 The Program tab contains the two following subpages: Program Main time program with all program steps. Remote configuratio

Pagina 387

6.5 761 Compact IC IC Net 2.3 87List of program instructions The following program instructions can be added to the time pro-gram on the Program s

Pagina 388

6 Devices IC Net 2.3 88 Remote configuration On the Remote configuration subtab user-defined remote com-mands can be defined, which can be inserted

Pagina 389

6.5 761 Compact IC IC Net 2.3 89 761 unit version Selection of instrument version: 761.0010 = 761 Compact IC without sup-pressor 761.0020 = 761 C