6.27 861 Advanced Compact IC
IC Net 2.3
245
861 unit version Selection of instrument version:
Suppressor Suppressor module:
Autostep with Automatic switching to the next position
each time the injection valve is switched ei-
ther to the Fill or to the Inject position.
Use peristaltic pump
If this option is disabled, the peristaltic
pump is not switched on automatically at
the start of a determination or with Startup
hardware
.
Use CO2 Suppressor (only displayed with version 2.861.0040)
If this option is disabled, the 853 CO
2
Sup-
pressor is not switched on automatically at
the start of a determination or with
Startup
hardware
.
Polarity Selection of the output signal polarity:
+ positive polarity (for anions)
negative polarity (for cations)
Links
The
Links tab of the 861 Advanced Compact IC is used for COM
port selection and settings (details see section 5.2.4 Links).
6.27.4 Hardware settings
Selection of the Hardware menu item of the 861 icon using the right
mouse button opens the hardware settings window consisting of
the three tabs
Hardware, Outputs and Alarm stops.
Hardware