6.23 766 IC Sample Processor
IC Net 2.3
181
Num Row number (read-only).
Task Program instruction (see below).
Parameter Parameter for program instruction (see be-
low).
Enabled Enable program start (a disabled program
is not started).
<Add> Add new program instruction.
<Delete> Delete selected program instruction.
List of program instructions
The following program instructions can be added to the time pro-
gram on the Program tab:
Instruction Parameter entry Meaning
Move sample, sample+1,
spec.#, 1...n
Move rack to place the specified
vessel at the needle.
Lift shift, work, rinse,
special,
0, 1...125 mm
Raise or lower the lift to the speci-
fied position.
Pump on, off, 1...999 s Switch on/off peristaltic pump at
the tower.
Scan "scan command" Scan remote interface until the
specified signal is received. In ad-
dition to the normal scan com-
mands (details see 766 Instruc-
tion for Use) special commands
defined on the Configuration tab
can be used.