6.27 861 Advanced Compact IC
IC Net 2.3
247
For a precise determination of the cell con-
stant, pump a calibration solution of known
conductivity through the IC system, observe
the displayed conductivity and change the
cell constant until the correct conductivity
value is displayed.
<Break times> Possibility for changing the break times for
injection valve Valve and suppressor mod-
ule Suppressor.
Outputs
The Outputs tab of the hardware settings window defines remote
output signals to be set automatically if specific events occur.
Event Events for automatic remote signal output:
Inject
Switching of the injection valve to the "IN-
JECT" position.
Fill
Switching of the injection valve to the "FILL"
position.
Suppressor step
Switching of the suppressor module to the
next position.
Pump stop (P max/min)
Pump stopped because pressure limits are
exceeded.
Hardware error
Hardware error detected at the 861 Ad-
vanced Compact IC (high-pressure pump,
injection valve, or suppressor not working
correctly).
Alarm leak detector
Leak detector has detected solvent in the
instruments interior.
Remote line
Set remote output lines 18.
Selection: 0 (line off, inactive, open)