6 Devices
IC Net 2.3
250
Controls
On the subtab Controls injection valve, peristaltic pump, high-
pressure pump, column heating and remote lines can be manu-
ally controlled. In addition, the parameters for the high-pressure
pump and column heating temperature can also be set:
Valve Injection valve
<Inject> Switch injection valve to "Inject" position.
<Fill> Switch injection valve to "Fill" position.
Peristaltic pump Peristaltic pump
<On> Start peristaltic pump (only enabled, with
versions 2.844.0020 or 2.844.120).
<Off> Stop peristaltic pump (only enabled, with
versions 2.844.0020 or 2.844.120).
IC pump High pressure pump
<On> Start high pressure pump.
<Off> Stop high pressure pump.
Flow, mL/min Set value for flow rate of IC pump.
Entry range:
0.20 ... 2.50 mL/min
Pressure max, MPa Value for maximum switch-off pressure for
high-pressure pump. Once it has been set
this value will be monitored, even if there is
no connection to the PC (units depend on
the settings made under IC NET / Global
preferences
).
Entry range:
0.0 ... 35.0 MPa
If the Flow is higher than 1.5 mL/min, the
defined Pressure max must not exceed