6.14 753 Suppressor Module
IC Net 2.3
149
First column Time at which program instruction is ap-
plied.
Entry range: 0.0 ... 999.9 min
If no time is entered, the program instruc-
tion is applied together with the last instruc-
tion with time entry.
Second column Program instruction (see below).
Third column Parameter for program instruction (see be-
low).
ENABLED Enable program start (a disabled program
is not started).
<Add> Add new program instruction.
<Delete> Delete selected program instruction.
<Verify> Test the time program (error messages are
displayed if program is wrong).
List of program instructions
The following program instructions can be added to the time pro-
gram on the Program page:
Instruction Parameter entry Meaning
Program RESET, END
Set program flag. The END flag
can be used to end a program,
especially if the program time
should be longer than the chro-
matogram
duration. Additional
steps after this flag are not al-
lowed. The
RESET flag is used to
reset the parameters to the sys-
tem startup values.
Pump on, off Switch on or off the pump drive.
Step Rotate suppressor to the next
position.
Links
The
Links tab of the 753 Suppressor Module is used for Event line
selection.