3 General settings for exploratory and determination mode
797 VA Computrace – Software
48
CV – Cyclic Voltammetry
General:
CV or Cyclic voltammetry is mainly used to investigate the re-
versibility of electrode processes and for kinetic studies. The CV
measurement mode can be set for the exploratory and determina-
tion mode by selecting CV - Cyclic voltammetry for the Mode pa-
rameter in the EXPLORATORY SPECIFICATION or WORKING
METHOD SPECIFICATIONS
window.
Description:
For cyclic voltammetry, the voltage is once or several times chan-
ged at a constant sweep rate to the end potential and then decrea-
sed at the same rate back to the start potential. The current i is
measured as a function of the voltage U. The curve registered in
the last cycle is stored and its peaks can be evaluated using linear,
polynomial, horizontal or exponential baselines.
i
an
U
i
cath
U
t
Start potential
First Vertex potential
Voltage step
Voltage
step time
Second Vertex potential