5 Determination mode
797 VA Computrace – Software
114
with the Calibration technique “MLAT Standard
addition for brighteners”.
The window displays date and time of the deter-
mination start and the title of the method used.
Date and time in the format
YYMMDDHHMM
(month-day-hour-minute) and the Sample ID are
used as default for the name of the determination
file to be saved automatically (e.g.
0706181712_sample.dth).
Sample ID [ 32 characters ; "sample" ]
Identification for sample. The expression defined
in the Determination tab is displayed and can be
changed if desired.
Volume intercept solution (mL) [> 0 mL ; 10 mL] [nur An-
zeige]
The volume of Intercept solution added is dis-
played.
Volume production bath (mL) [ > 0 mL ; 10 mL ]
The volume of “sample bath solution” added (see
Production bath solution, section 6.4) is displayed
and can be changed if desired.
Cell volume (mL) [ > 0 mL ; 10 mL ]
Total volume of solution (sample + auxiliary solu-
tion, e.g. buffer (manually or pre-dosed with a
Dosing Device)) in the measuring vessel at the
start of the determination. The calculated sample
concentrations Conc. refer to this cell volume. It
can be changed if desired.
Add the sample solution into the measuring vessel
and press <OK>.
PLACE BATH SOLUTION
This window appears in the Electroplating Bath VA
with the
Calibration techniques "LAT Standard ad-
dition for brighteners" and "RC Sample with re-
sponse curve".