Metrohm 797 VA Computrace Manual de usuario Pagina 211

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 295
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 210
7.2 Dosing Device control
797 VA Computrace – Software
199
7.2 Dosing Device control
Dosing Device control selection
MAIN WINDOW / Utility / Dosino control
Start manual control of the Dosing Devices (possi-
ble: 700/800 Dosino , 685/805 Dosimat) con-
nected to the 797 VA Computrace stand or the
846 Dosing Interface.
Dosino control window
The DOSINO CONTROL window serves for manual control of the
Dosing Devices connected to the 797 VA Computrace stand (under
Dosing Processor) or the 846 Dosing Interface (under Dosing Inter-
face
).
Dosino 1..7 (x mL)
3 Dosing Devices can be connected to the 797 VA
Computrace stand, 4 more to a 846 Dosing In-
terface. Each can be controlled individually.
Status [ read only ]
Display of the current Dosing Device status.
Total dosed volume (mL) [ > 0.01 mL ; 0 mL ]
Display of current volume dispensed since last fil-
ling (first value) and accumulated volume if the
display is not reset after filling (value in brackets).
Vista de pagina 210
1 2 ... 206 207 208 209 210 211 212 213 214 215 216 ... 294 295

Comentarios a estos manuales

Sin comentarios