Metrohm 797 VA Computrace Manual de usuario Pagina 245

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 295
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 244
8.6 Analyze Electroplating Bath Solutions
797 VA Computrace – Software
233
Note: To make sure that always the latest intercept file is taken
for the calculation, the name of the intercept file defined for the
parameter Intercept determination, must match with the pa-
rameter Sample identifier on the Determination tab with Calibra-
tion
technique "LAT Record intercept value".
Note: To make sure that always the latest intercept file is taken
for the calculation, the path of the intercept file defined for the
parameter Intercept determination, must match with the path
defined for the parameter Data folder (for the currently logged-
in user) on the tab User Directories of the window USER RIGHTS.
Sample table
In the Sample table, itemize alternately (based on the arrangement
on the sample rack) a method with
Calibration technique "LAT Re-
cord intercept value" (as first) and a method with Calibration tech-
nique "LAT Standard addition for brighteners":
Start determination.
Vista de pagina 244
1 2 ... 240 241 242 243 244 245 246 247 248 249 250 ... 294 295

Comentarios a estos manuales

Sin comentarios