3.2 VA measurement modes
757 VA Computrace – Software
39
CV – Cyclic Voltammetry
General:
CV or Cyclic voltammetry is mainly used to investigate the re-
versibility of electrode processes and for kinetic studies. The CV
measurement mode can be set for the exploratory and determi-
nation mode by selecting
CV - Cyclic voltammetry
for the
Mode
pa-
rameter in the
EXPLORATORY SPECIFICATION
or
WORKING
METHOD SPECIFICATIONS
window.
Description:
For cyclic voltammetry, the voltage is once or several times
changed at a rapid but constant sweep rate to the end potential
and then decreased at the same rate back to the start potential.
The current i is measured as a function of the voltage U. The
curve registered in the last cycle is stored and its peaks can be
evaluated using linear, polynomial or exponential baselines.
i
an
U
i
cath
U
t
Start potential
Voltage step
Voltage
step time
End potential
t (
i
)