7.9 Diagnostic procedures
Check the purging
1. Connect the inert gas to the 757 VA Computrace Stand (see
Hardware Manual).
2. Make sure that the inert gas pressure is 1 ± 0.2 bar.
3. Add 20 mL ultrapure water to the empty measuring vessel at
the 757 VA Computrace Stand.
4. Click on or
MAIN WINDOW
/ Utility / Computrace control
to
open the
COMPUTRACE CONTROL
window.
5. Select
HMDE
and click on .
6. Make sure that inert gas bubbles are purging through the so-
lution.
Check the stirring
1. Click on or
MAIN WINDOW
/ Utility / Computrace control
to
open the
COMPUTRACE CONTROL
window.
2. Select
HMDE
and click on .
3. Change the rotational speed by clicking on the buttons of
the
RDE/stirrer speed
field.
Check the MME
1. Install the MME at the 757 VA Computrace Stand (see Hard-
ware Manual).
2. Make sure that the inert gas pressure is 1 ± 0.2 bar.
3. Add 10 mL ultrapure water to the empty measuring vessel at
the 757 VA Computrace Stand.
4. Add 0.25 mL potassium chloride c(KCl) = 1 g/L (Metrohm
No. 6.2308.020) to the measuring vessel.
5. Click on or
MAIN WINDOW
/ Utility / Computrace control
to
open the
COMPUTRACE CONTROL
window.
6. Select
DME
and click on .
7. Mercury flows out of the capillary. The typical natural drop
time is in the range of 1 - 5 sec. If the natural drop time is
smaller than 0.5 sec, change the capillary and/or needle at
the MME (see Hardware Manual).