4 Troubleshooting
780/781 pH/Ion Meter - Operation via RS232 27
4 Troubleshooting
Problem Questions for remedical action
No characters can be
received on a
connected printer.
Are the instruments switched on and cables plugged in
correctly?
Is the printer set to ”on-line”?
Are baud rate, data bit and parity the same on both instruments?
Is the handshake set properly?
If everything seems to be ok, try to print a report with the key
sequence <PRINT><SMPL DATA><ENTER>. If this report is
printed out correctly, check if reports are defined in key <DEF>.
No data transmission
and the display of the
pH/Ion Meter shows
an error message.
error 42: Transmission error. Is the printer set to "on-line"? Is the
connection cable properly wired?
error 43: Data output of the pH/Ion Meter disabled for longer
than 6 s by XOFF.
error 36-39: Receive error. Are the RS settings the same on both
devices?
The received
characters are
garbled.
Are the RS settings the same on both devices?
Has the correct printer been selected?
Data transfer has been interrupted on the hardware side during
the printout of a curve. Re-establish connections and switch
printer off/on.
Wrong line spacing. The printer does not emulate completely the preset mode. Usually
these problems arise with the IBM mode. Set the printer to a
different mode (e.g. Epson).
Printout of curve is not
ok. Other reports are
printed ok.
Handshake is necessary for the printout of curves.
Is your cable correctly wired? (The DTR of the printer has to be
connected to the CTS of the pH/Ion Meter.)
Set "HWs" for the handshake of the pH/Ion Meter. Configure the
printer such that its DTR is set (possibly with DIP switches).