2 Remote control commands
780/781 pH/Ion Meter - Operation via RS232 15
&UserMeth
Object Description Input range
&UserMeth Method memory
: ├ .FreeMemory Memory available read only
├ .Recall Load method $G
│ └ .Name Method name 8 ASCII char.
├ .Store Save method $G
│ └ .Name Method name 8 ASCII char.
├ .Delete Delete method $G
│ └ .Name Method name 8 ASCII char.
├ .DeleteAll Delete all methods $G
└ .List List of methods
│ .1 Method 1
│ ├ .Name Method name read only
│ ├ .Mode Mode read only
│ ├ .Bytes Method size in bytes read only
│ └ .Checksum Checksum of method read only
└ .n for each method
&Config
Object Description Input range
&Config Instrument configuration
: ├ .Report Report configuration
│ ├ .Id1 Identification 1 for report header 16 ASCII char.
│ ├ .Id2 Identification 2 for report header 16 ASCII char.
│ ├ .IdReport Ids for report header ON, OFF
│ ├ .ReDevName Print device name ON, OFF
│ ├ .ReDateTime Print date and time ON, OFF
│ ├ .ReMethod Print method name ON, OFF
│ ├ .ReElId Print electrode identification ON, OFF
│ ├ .ReVisum Print signature line ON, OFF
│ └ .ReLineFeed Paper feed after report printing 0,1,2,3...999, form
├ .PrintMeasVal Print measured values
│ ├ .PrintCrit Criterion for printing measured values immediate, time, drift, plot, OFF
│ ├ .DateTime Print date & time with measured value ON, OFF
│ ├ .PrintHead Print header with measured value once, always, OFF
│ ├ .CalRep Print calibration report with meas. val. ON, OFF
│ ├ .Time
│ │ ├ .Interval Time interval for printing meas. val. [s] 0.1...4.0...99999.0
│ │ └ .StopTime Stop time for printing [s] 1...999999, OFF
│ └ .Plot
│ ├ .Interval Time interv. for printing meas. point [s] 0.1...4.0...99999.0
│ ├ .TimeScale Time scale for plot [s/cm] 5, 10, 30, 60, 120, 180... 99960
│ ├ .TLabel Clock time or seconds on time scale abs, rel
│ └ .StopTime Stop time for printing [s] 1...999999, OFF
├ .StoreMeasVal Store measured values
│ ├ .StoreCrit Criterion for storing measured values immediate, time, drift, OFF
│ └ .Time
│ ├ .StoreInterval Time interv. for storing meas. val. [s] 0.1...4.0...99999.0
│ └ .StoreStopTime Stop time for storing [s] 1...999999, OFF
Comentarios a estos manuales