Metrohm 781 pH/Ion Meter: Operation via RS-232 Manual de usuario Pagina 18

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 29
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 17
1.4 Error messages
16 780/781 pH/Ion Meter - Operation via RS232
.Aux Miscellaneous
.RunNo Current run number 0...999, OFF
.LastDigit Last digit of display ON, OFF
.Language Dialog language english, deutsch, français, español
.Display Display type positiv, negativ
.ScreenSave Switch LCD display off after [min] 1,2...999, OFF
.Set Sets date and time $G
.Date Current date YYYY-MM-DD
.Time Current time hh:mm:ss
.TimeZone Text for time zone 12 ASCII char.
.TSensor Type of temperature sensor PT1000, NTC
.NTCFactor Factors for NTC sensor
.NTC25 R(25°C) value 10000...30000...100000
.TSlope Slope, 'B value' 1000...4100...9999
.TempUnit Temperature unit °C or °F C, F
.DevName Device label 12 ASCII char.
.Beeper Number of acoustic signals 1,2,3, OFF
.Prog Program version number read only
.Monitor Monitoring functions
.Valid Validation monitoring
.VState Status of validation monitoring ON, OFF
.VInterval Time interval for validation [d] 1...365...9999
.VCounter Time counter [d] 0...9999
.ClearCount Clears the counter above $G
.Service Monitoring of Metrohm service $G
.SState Status of service monitoring ON, OFF
.SDate Date of next service YYYY-MM-DD
.SysTestReport Printing of system test report ON, OFF
.Periph Selection of peripheral units
.CharSet Selection of character set IBM, Epson, Seiko, Citizen, HP
.Dosimat Selection of Dosimat type 665, 725, 765, 776
.SelStirrer Selection of stirrer type 7xx, 8xx
.PcKeyb Type of keyboard US, deutsch, français, español,
schweiz.
.Barcode Input of barcode reader input, id1, id2
.RSset Settings RS232
.Baud Baud rate 38400, 19200, 9600, 4800,
2400, 1200, 600, 300
.DataBit Number of data bits 7, 8
.StopBit Number of stop bits 1, 2
.Parity Parity none, odd, even
.Ha ndsh Handsh ake HWs, SWchar, SWline, none
&Info
Object Description Input range
&Info Current data
: .Report Transmission of formatted reports $G
.Select Selection of report type user memory, calib short, calib full,
config, param, el.test, result short,
result full, mv memory, all
.pHCalData pH calibration data
.ElectrodeId Identification of calibrated electrode read only
.MethodId Name of calibration method read only
.Slope Slope [%] 0.1...100.0...999.9
.pH0 Electrode zero point -99.999...7.0...99.999
.CalTemp Calibration temperature read only
Vista de pagina 17