2 Remote control commands
780/781 pH/Ion Meter - Operation via RS232 13
│ │ ├ .LeftMargin Left margin of temperature scale [°C] -999.9...0.0...999.9
│ │ └ .RightMargin Right margin of temp. scale [°C] -999.9...100.0...999.9
│ └ .Presel Preselections
│ └ .IReq Request of identifications after start id1, id1 & id2, OFF
│
└ .Conc Measuring mode Conc (concentration) (781 only)
├ .MeasType Selection of measuring type direct, std.add, smpl.add
├ .IonPara Ion/Concentration $G, $S
│ ├ .Ion
│ │ ├ .Select Selection of ion type Ag(+1), BF4(–1), Br(–1), Ca(+2),
│ │ │ Cd(+2), Cl(–1), CN(–1), Cu(+2),
│ │ │ F(–1), I(–1), K(+1), Na(+1), NH4(+1),
│ │ │ NO2(–1), NO3(–1), Pb(+2), S(–2),
│ │ │ SCN(–1), SO4(-2), own
│ │ └ .Own
│ │ ├ .Name Name of ion 7 ASCII char.
│ │ └ .Charge Charge of ion -9...-1, 1...9
│ └ .Unit
│ ├ .Select Unit for concentration mol/L, %, ppm, g/L, mg/L, µg/L,
│ │ mEq/L, own
│
└ .Own String for own unit 5 ASCII char.
│
├ .MeasPara Measuring parameters
│ ├ .ElectrodeId Electrode identification 12 ASCII char.
│ ├ .Drift Drift f. meas. val. acquisition [mV/min] 0.1...1.0...999.9, OFF
│ ├ .Temperature Measuring temperature [°C] -999.9...25.0...999.9
│ ├ .MethodId Method name 8 ASCII char. (read only)
│ ├ .Delta Delta measurement
│ │ ├ .Status Status of delta measurement ON, OFF
│ │ └ .Reference Reference for delta measurement -1.00E+30...0.00...1.00E+30
│ └ .Stirrer
│ ├ .Status Status of stirrer ON, OFF, control
│ ├ .Rate Stirring rate 1...5...15
│ ├ .PreStirTime Pause before stirring [s] 0...99999
│ ├ .StirTime Stirring time [s] 0...99999
│ └ .PostStirTime Pause after stirring [s] 0...99999
├ .CalcPara Calculation parameters
│ ├ .SmplSize Sample size [mL] or [g] 0.0001...1.0...99999.9, OFF
│ ├ .VTotal Total starting volume [mL] 0.001...100.0...9999.9
│ ├ .Factor Factor (result multiplier) -1.0E+30...1.0...1.0E+30
│ └ .SmplUnit Unit for sample size mL, g
├ .Direct Parameters for direct measurement
│ ├ .Cal $G, $S
│ ├ .CalPara Calibration parameters
│ │ ├ .CalTemp Calibration temperature [°C] 0.0...25.0...99.0
│ │ ├ .Drift Drift for meas. val. acqu. [mV/min] 0.1...0.5...9.9
│ │ ├ .Report Automatic printout of calibration report short, full, OFF
│ │ ├ .CalInterval Calibration interval [h] 1...999, OFF
│ │ ├ .NumberStd Number of standards 1, 2...19
│ │ ├ .Type Type of standard handling manual, auto
│ │ ├ .Manual Manual calibration
│ │ │ ├ .1
│ │ │ │ └ .Conc Concentration of standard 1 1.0E-30...0.01...1.0E+30
│ │ │ ├ .2
│ │ │ │ └ .Conc Concentration of standard 2 1.0E-30...0.01...1.0E+30
│ │ │ └ up to 19
│ │ └ .Auto Definition of own buffers
│ │ ├ .CMin Minimal concentration of standard 1.0E-30...0.1...1.0E+30