2 Remote control commands
780/781 pH/Ion Meter - Operation via RS232 13
.LeftMargin Left margin of temperature scale [°C] -999.9...0.0...999.9
.RightMargin Right margin of temp. scale [°C] -999.9...100.0...999.9
.Presel Preselections
.IReq Request of identifications after start id1, id1 & id2, OFF
.Conc Measuring mode Conc (concentration) (781 only)
.MeasType Selection of measuring type direct, std.add, smpl.add
.IonPara Ion/Concentration $G, $S
.Ion
.Select Selection of ion type Ag(+1), BF4(–1), Br(–1), Ca(+2),
Cd(+2), Cl(–1), CN(–1), Cu(+2),
F(–1), I(–1), K(+1), Na(+1), NH4(+1),
NO2(–1), NO3(–1), Pb(+2), S(–2),
SCN(–1), SO4(-2), own
.Own
.Name Name of ion 7 ASCII char.
.Charge Charge of ion -9...-1, 1...9
.Unit
.Select Unit for concentration mol/L, %, ppm, g/L, mg/L, µg/L,
mEq/L, own
.Own String for own unit 5 ASCII char.
.MeasPara Measuring parameters
.ElectrodeId Electrode identification 12 ASCII char.
.Drift Drift f. meas. val. acquisition [mV/min] 0.1...1.0...999.9, OFF
.Temperature Measuring temperature [°C] -999.9...25.0...999.9
.MethodId Method name 8 ASCII char. (read only)
.Delta Delta measurement
.Status Status of delta measurement ON, OFF
.Reference Reference for delta measurement -1.00E+30...0.00...1.00E+30
.Stirrer
.Status Status of stirrer ON, OFF, control
.Rate Stirring rate 1...5...15
.PreStirTime Pause before stirring [s] 0...99999
.StirTime Stirring time [s] 0...99999
.PostStirTime Pause after stirring [s] 0...99999
.CalcPara Calculation parameters
.SmplSize Sample size [mL] or [g] 0.0001...1.0...99999.9, OFF
.VTotal Total starting volume [mL] 0.001...100.0...9999.9
.Factor Factor (result multiplier) -1.0E+30...1.0...1.0E+30
.SmplUnit Unit for sample size mL, g
.Direct Parameters for direct measurement
.Cal $G, $S
.CalPara Calibration parameters
.CalTemp Calibration temperature [°C] 0.0...25.0...99.0
.Drift Drift for meas. val. acqu. [mV/min] 0.1...0.5...9.9
.Report Automatic printout of calibration report short, full, OFF
.CalInterval Calibration interval [h] 1...999, OFF
.NumberStd Number of standards 1, 2...19
.Type Type of standard handling manual, auto
.Manual Manual calibration
.1
.Conc Concentration of standard 1 1.0E-30...0.01...1.0E+30
.2
.Conc Concentration of standard 2 1.0E-30...0.01...1.0E+30
up to 19
.Auto Definition of own buffers
.CMin Minimal concentration of standard 1.0E-30...0.1...1.0E+30