6 Devices
IC Net 2.2
70
Cell constant Cell constant of the conductivity cell for cor-
rect display of the absolute conductivity.
Enter the cell constant printed on the detec-
tor block into this field.
Range: 0.1 ... 99.9 /cm
For a precise determination of the cell con-
stant, pump a calibration solution of known
conductivity through the IC system, observe
the displayed conductivity and change the
cell constant until the correct conductivity
value is displayed.
<Break times> Possibility for changing the break times for
injection valve Valve and suppressor mod-
ule
Suppressor.
These values should only be changed af-
ter consulting the Metrohm Service.
Outputs
The Outputs tab of the hardware settings window defines remote
output signals to be set automatically if specific events occur.
Event Events for automatic remote signal output:
Inject
Switching of the injection valve to the "IN-
JECT" position.
Fill
Switching of the injection valve to the "FILL"
position.
Suppressor step
Switching of the suppressor module to the
next position.