Metrohm IC Net 2.2 Manual de usuario Pagina 185

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 249
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 184
7.6 Report output
IC Net 2.2
177
4 11.19 5.0 9.933e-03 0.00 0.000 0 bromide
5 14.60 5.0 7.912e-03 7.56 0.000 S 0 nitrate
6 16.80 5.0 1.556e-02 0.00 0.000 0 phosphate
7 21.26 5.0 6.021e-03 5.57 0.000 0 sulphate
Calibration results Calibration results of the component win-
dow. For each component a new page with
calibration results and calibration curve is
put out.
CALIBRATION OF COMPONENT fluoride
Method: Multicalibration.mtw
Equation: Q = - 7.48163·A^2 + 0.00262326·A + 0.0233088
RSD: 0.691 %
Correlation coefficient: 0.999277
K3 = 0 K2 = -7.48163e-08 K1 = 0.00262326 K0 = 0.0233088
Base: Area
Ref.channel: ch1
ISTD:
Formula: Y=K2·X^2+K1·X+K0
Weight: 1
Level Height Area Conc. Volume Retention Used File
1 6.46 173 2 1 10.8 Yes g8140800.chw
2 34 897 10 1 10.8 Yes g8140823.chw
3 90.9 2.36e+03 25 1 10.8 Yes g8140858.chw
4 198 5.04e+03 50 1 10.8 Yes g8140929.chw
Channel table Parameters of the data acquisition channel
used.
CHANNELS TABLE
ch1
No Name Units Input Minimum Zero Maximum Range Coefficient Noise Shift
1 ch1 mV 1 -8388607 0 8388607 2.500e+03 2.980e-04 4.15 0
# Name/Units Noise RMS PeakToPeak Drift/hour
1 ch1 4.2 29976.5 330509.6 3585.801
mV 0.00124 8.93 98.5 1.07
Name
Channel’s name. This text appears as
Channel label in the chromatogram.
Units User-defined units of detector response.
Input Number of data acquisition channel.
Minimum Minimum value of the linear range of the AD
converter (in ADC conversion units). This
Vista de pagina 184
1 2 ... 180 181 182 183 184 185 186 187 188 189 190 ... 248 249

Comentarios a estos manuales

Sin comentarios