8 Chromatograms
IC Net 2.2
192
Represent raw data as Output of raw data:
Default units (float) Y values are exported in default units for the
channel.
ADC response (integer)
Y values are exported in bits.
Include retention time
In seconds (float) X values are exported in seconds.
As point number X values are exported as point number.
8.2.6 Import chromatogram
IC NET / File / Import / Chromatogram
Import chromatogram of the following formats:
*.chm Chromatogram files of the «Chrom&Spec»
program (DOS version).
*._rd Chromatogram files of the «714 IC Metro-
data» program (DOS version). After selection
of the file to be imported you are asked to
load a method.
*.chr Chromatogram files of the «EnviroChrom»
program.
*.dar Chromatogram files of the «AtomChrom»
program.
LongInteger (*.*) Chromatogram files in Long Integer format.
IC NET / File / Import / AIA file
Import chromatogram in the AIA format (Analytical Instrument As-
sociation) from a CDF file (
*.cdf).
IC NET / File / Import / Raw data from txt
Import chromatographic raw data from a text file *.txt (ASCII for-
mat). This is only possible if an appropriate method has been
opened.