6 Devices
IC Net 2.1
54
List of program instructions
The following program instructions can be added to the time pro-
gram on the
Program
subpage:
Instruction Parameter entry Meaning
Remote 0, 1, *
Set remote output lines 1...8 to
the desired values. For entry of
the first value, enter
1
,
0
or
*
. For
entry of the other values, move
the cursor in front of the value to
be changed and enter
1
,
0
or
*
.
Flag reset, end
Set program flag. The
end
flag
can be used to end a program,
especially if the program time
should be longer than the chro-
matogram
duration
. Additional
steps after this flag are not al-
lowed. The
reset
flag is used to
reset the parameters to the sys-
tem startup values.
Range 100, 200, 5000 P
PP
PS/cm
1, 2, 5, 10 mS/cm
Set measuring range to the se-
lected value.
FS 0.05 P
PP
PS/cm ...
10 mS/cm
Set full scale range to the se-
lected value (depends on
Range
setting).
Zero on, off
Switch on or off the auto-zero
function.
Polarity +,
Reset the polarity of the output
signal.
Mark
Trigger marking signal.
ValveA fill, inject
Switch injection valve A to "
inject
"
or "
fill
" position.
ValveB fill, inject
Switch injection valve B to "
inject
"
or "
fill
" position.
Suppressor
Switch suppressor module to the
next position.
PumpRS on, off
Switch on or off the high-pressure
pump connected to the 732 IC
Detector.
Flow 0.01 ... 5 mL/min
Set flow rate of the high-pressure
pump to the desired value.