Metrohm IC Net 2.1 Manual de usuario Pagina 198

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 228
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 197
8 Chromatograms
IC Net 2.1
190
8.6.4 Data compression
IC NET / Process / More / Compress
Selection of this menu item opens the
Raw data compression
win-
dow for compression of the raw data of the active chromatogram
by summing up several neighboring points.
Divisor
Coefficient of compression (the number of
data points is reduced by this value). The
default value of the compression coefficient
is calculated so that the width at half-height
of the narrowest peak will be digitized by at
least 30 points. If the user enters a coeffi-
cient of compression that exceeds the
value offered by the software, integration
precision may decrease.
8.6.5 Invert chromatogram
IC NET / Process / More / Invert!
This menu item inverts the response curves for all channels of the
chromatogram so that negative peaks become positive and vice
versa (useful for chromatograms with wrong input polarity).
Vista de pagina 197
1 2 ... 193 194 195 196 197 198 199 200 201 202 203 ... 227 228

Comentarios a estos manuales

Sin comentarios