7 Methods
IC Net 2.1
156
Level 1...Level N
Contains concentrations (or other chosen
values) of components for corresponding
Calibration level
.
<Add>
Add a new
Calibration level
to the
concentration table. The following window
appears:
Create calibration level
Number of calibration level to add.
Fill level with concentration
Concentration to be filled in for all compo-
nents.
Copy concentrations from level
Number of calibration level from which con-
centrations are copied.
Calibrate immediately
Recalibration with new calibration levels.
<Delete>
Delete the current calibration level (where
cursor is situated) from the concentration
table.
<Calibrate>
Calibrate the current chromatogram with
the selected calibration level. The following
window appears:
Level
Calibration level to be used for recalibra-
tion.