6 Devices
IC Net 2.1
50
6.3.3 732 IC Detector window
732 icon / Open
The
732 IC Detector
window for parameter settings is opened by
selecting this menu option with the right mouse button or by dou-
ble-clicking the
732 icon
in the
SYSTEM
window. It consists of the
four tabs
Display
,
Method parameter
,
Program
, and
Links
.
Display
The
Display
tab of the
732 IC Detector
window is only available for a
connected system. It displays the current measurement values
and parameters settings of the 732 IC Detector.
Conductivity
Live display of measured conductivity
(auto-zero value) in PS/cm. This field is
available for the
WATCH WINDOW
. The
watch window color is changed to the
Out
of range
color if the measured value is more
than 80% outside the
Full Scale
range
.
abs. Cond.
Live display of absolute conductivity value
in PS/cm. This field is available for the
WATCH WINDOW
. The watch window color is
changed to the
Out of range
color if the
measured value is bigger or equal to 110%
of
Range.
Ion
Display of ion settings:
Anion
:
polarity = +, temperature coeff. = 2.5
Cation
:
polarity = –, temperature coeff. = 1.5