Metrohm 827 pH lab Manual de usuario Pagina 46

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 82
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 45
7.1 Reports
38 826/827 pH meter, Instructions for Use
7.1.6 Parameter report
The complete set of parameters of the mode pH is printed out with a
parameter report. A parameter report can't be printed for the measuring
modes U (mV) and T (°C) because these modes don't have any edita-
ble parameters.
7.1.7 Measured values memory report
Up to 200 measured values can be stored in the menu << store >>
(see Section 7.2). If you want to print out all the stored data then you
can produce a measured values memory report.
The 'id' will only be included in the report if an identification has been
entered under configuration/report.
7.1.8 PC/LIMS report
The PC/LIMS report is a machine readable report of all the stored
measured values. It can be sent to a PC for storage (e. g. Metrohm Ve-
suv
©
3.0 for Windows
TM
).
The PC/LIMS report has got the following structure:
Start of the data set:
$S
Report identification:
'md
Device identification:
827 pH lab
Serial number:
01106
Program version:
5.827.0010
Device identification (
dev.label
):
lab 1a
No. of measured value:
#1
Measuring quantity 1:
pH/U/T
Value:
4.477/165.3/24.3
Unit:
--/mV/°C NTC
Measuring quantity 2 (temperature):
T/--/--
Value:
23.8/--/--
Temperature unit, sensor type:
°C NTC/--/--
Remark (id):
sample 1
Date, time:
2005-01-31 17:05:39
Start of the data set:
$E
--------------
<< report >>
param
<< report >>
mval
<< report >>
pc/lims
Vista de pagina 45