4.2 pH calibration
18 826/827 pH meter, Instructions for Use
3 Result
• Wait for potential measurement, the result will be displayed:
• After 30 s the pH meter will switch back automatically to the
measurement display. This can also be done immediately with
<OK> or <QUIT>.
3 Poor calibration data
• If the calibration data are outside the calibration parameters
defined as the limits (see Section 6.1.3) then a corresponding
message will be shown:
• You can still accept the calibration data with <OK>, or reject
it with <QUIT>.
4 Show calibration data
You can now display the calibration data under param/
cal.data.
5 Print calibration report
• Change to the menu << report >> with the arrow keys <
>
or <
>.
• Press <OK>.
• Select the report cal with the arrow keys <
> or <
> and
start the printing with <OK>.