6.1 pH measurement (pH mode)
32 826/827 pH meter, Instructions for Use
The other stored buffer tables are described in Section 9.3. During the
calibration the pH meter shows the type and temperature-specific pH
value of the recognized buffer. For temperatures within the 5 °C steps
the pH value is linearly interpolated.
buff.type: special
If you would like to use different buffers from those described,
then you require the exact pH value of each buffer at the tempera-
ture used for calibration. These pH values can be entered here.
During the calibration they can again be adapted before each
measurement.
A general rule is that you can only calibrate in defined temperature
ranges, as otherwise the following error message appears:
buffer not defined
6.1.3 Limits
param
:
cal.limits
: lo.lim.slope
up lim.slope
lo.lim.pH(0)
up lim.pH(0)
The most important calibration data such as slope and pH(0) must be
within the limits defined here in order to be adopted automatically.
lo.lim.slope 1.0...95.0...999.9 %
up lim.slope 1.0...103.0...999.9 %
lo.lim.pH(0) 0.00...6.40...99.99
up lim.pH(0) 0.00...8.00...99.99
If one of the values of the calibration data lies outside these limits then
at the end of the calibration the following message is displayed:
cal. out of limits
You can accept these calibration data nevertheless with <OK> or you
can reject them with <QUIT>.