Metrohm IC Net 2.0 Manual de usuario Pagina 45

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 196
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 44
5.2 762 IC Interface
IC Net 2.0
37
Event command (2
nd
column)
Setting the event output lines 17.
Selection:
0 (line off, inactive, open)
1 (line on, active, 0 V)
p (pulse, pulse length 150 ms)
* (leave line in current status)
For entry of the first value, enter 1, 0, p or *.
For entry of the other values, move the cur-
sor in front of the value to be changed and
enter 1, 0, p or *.
<Add> Add new remote command.
<Delete> Delete selected remote command.
<Activate> Activate the defined remote commands for
insertion into the time program.
Events setup
The Events setup tab of the 762 System [#] window contains the
startup values for the seven event output lines of system 1 or
system 2 of the 762 IC Interface. These startup values are sent
each time the system is connected or a determination is started.
Each event output line can be set to the following values:
0 Line off, inactive (open)
1 Line on, active (0 V)
For technical information see 762 Instructions for Use.