5 Troubleshooting
856 Conductivity Module
■■■■■■■■
25
5 Troubleshooting
5.1 General
Problem Cause Remedy
Unstable measuring
signal - value near
zero.
The plug is plugged in
incorrectly.
The plug must be plugged in according to the
manual (dot to line).
Measured values dif-
fer strongly from
one another.
The cuvette is missing. Screw the cuvette on to the 5-ring conductivity
measuring cell. Measure only with cuvette!
The measurement
signal is influenced
by the movement of
the 5-ring conductiv-
ity measuring cell
and/or its position in
the measuring ves-
sel.
The cuvette is missing. Screw the cuvette on to the 5-ring conductivity
measuring cell. Measure only with cuvette!