Metrohm Inline-Ultrafiltration Manual de usuario Pagina 27

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 31
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 26
■■■■■■■■■■■■■■■■■■■■■■
4 Operation and maintenance
IC equipment: Inline-Ultrafiltration
■■■■■■■■
21
Tighten the PTFE capillary (6.1803.060) to the inlet labeled IN of
the ultrafiltration cell's lower chamber using a PVDF pressure
screw (6.2744.000).
Tighten the PTFE capillary (6.1803.080) to the outlet labeled OUT
of the ultrafiltration cell's lower chamber using a PVDF pressure
screw (6.2744.000).
Tighten the PTFE capillary (6.1803.050) to the outlet labeled OUT
of the ultrafiltration cell's upper chamber using a PVDF pressure
screw (6.2744.000).
4
Inserting the ultrafiltration cell into the holder
Place the ultrafiltration cell back in the holder (see Chapter 3.3,
page 14).
5
Deaerating the ultrafiltration cell
Carry out all steps contained in the instructions Rinsing the ultrafiltra-
tion cell on page 16.