4 Operation and maintenance
IC equipment: Inline-Ultrafiltration
■■■■■■■■
21
■
Tighten the PTFE capillary (6.1803.060) to the inlet labeled IN of
the ultrafiltration cell's lower chamber using a PVDF pressure
screw (6.2744.000).
■ Tighten the PTFE capillary (6.1803.080) to the outlet labeled OUT
of the ultrafiltration cell's lower chamber using a PVDF pressure
screw (6.2744.000).
■ Tighten the PTFE capillary (6.1803.050) to the outlet labeled OUT
of the ultrafiltration cell's upper chamber using a PVDF pressure
screw (6.2744.000).
4
Inserting the ultrafiltration cell into the holder
■ Place the ultrafiltration cell back in the holder (see Chapter 3.3,
page 14).
5
Deaerating the ultrafiltration cell
Carry out all steps contained in the instructions Rinsing the ultrafiltra-
tion cell on page 16.