3.2 Connecting the dialysis cell
■■■■■■■■■■■■■■■■■■■■■■
12
■■■■■■■■
IC equipment: Dialysis
■ Tighten a PTFE capillary, 0.5 mm ID / 1 m (6.1803.040), to the
outlet of the pump tubing with black stoppers (6.1826.340) using
a pressure screw (6.2744.070).
Tighten the other end of the PTFE capillary to the sample inlet of
the dialysis cell (3-4) using a pressure screw (6.2744.000).
4
Connecting the sample outlet
■ Tighten a PTFE capillary, 0.5 mm ID / 1 m (6.1803.040), to the
sample outlet of the dialysis cell (3-3) using a pressure screw
(6.2744.000).
Tighten the other end of the PTFE capillary to the inlet of the sec-
ond pump tubing with black stoppers (6.1826.340) using a pres-
sure screw (6.2744.070).
■ Tighten another PTFE capillary, 0.5 mm ID / 1 m (6.1803.040), to
the outlet of the second pump tubing with black stoppers
(6.1826.340) using a pressure screw (6.2744.070).
Tighten the other end of the PTFE capillary to the waste container.
5
Connecting the inlet of the acceptor solution