3 Installation
IC equipment: Dialysis
■■■■■■■■
11
The acceptor solution is pumped with two pump tubings with yellow/
orange stoppers (6.1826.320). Additional information on the tubing con-
figuration of the peristaltic pump can be found in the chapter "Installing
the peristaltic pump" in the manual for your ion chromatograph.
1
Preparing two pump tubings for the sample
Use the two pump tubings with black stoppers (6.1826.340) for con-
veying the sample. Proceed as follows for each tubing:
Attach the coupling olive/UNF 10/32 (6.2744.034) to the inlet.
Attach the pump tubing connection with locking nut
(6.2744.160) to the outlet (see chapter "Installing the peristaltic
pump" in the manual for the ion chromatograph or in the man-
ual for the Sample Processor).
2
Preparing two pump tubings for the acceptor solution
Use the two pump tubings with orange/yellow stoppers (6.1826.320)
for conveying the acceptor solution. Proceed as follows for each tub-
ing:
Attach the coupling olive/UNF 10/32 (6.2744.034) to the inlet.
Attach the pump tubing connection with locking nut
(6.2744.160) to the outlet (see chapter "Installing the peristaltic
pump" in the manual for the ion chromatograph or in the man-
ual for the Sample Processor).
3
Connecting the sample inlet
Tighten the PEEK capillary, 0.5 mm ID / 70 cm (6.1831.160), to
the inlet of the pump tubing with black stoppers (6.1826.340)
using a pressure screw (6.2744.070).
Connect the other end of the PEEK capillary to the needle of the
Sample Processor (see manual for the Sample Processor).